... ... @@ -82,16 +82,72 @@
82 82  
83 83  * **Tape:**
84 84  ** Click on the 'main tape' icon
85 -** Change the velocity by typing anything from 0.3-3.0 mm/s
85 +** Change the velocity by typing anything from **0.3-3.0 mm/s**
86 86  ** Common velocity is 1.0-1.5 mm/s
87 +** kamzik displays the used tape in m and %. At 98% the tape will automatically stop since the rolls need to be exchanged at this point.
87 87  ** [[image:attach:kamzik_tape.PNG||height="400"]]
88 88  
89 89  \\
90 90  
92 +* **Sample nozzle:**
93 +** Click an the 'nozzle' icon.
94 +** (((
95 +The nozzle position can be adjusted in xyz by clicking on the arrows up and down. Step size can be varied, 1mm steps are rather big and should be used carefully, for y-movement always use a stepsize <= 0.5mm.
96 +The absolute position is **not calibrated** and therefore changes after each restart of kamzik.
97 +\\Adjust the nozzle position visually until it nearly touches the tape and roughly meets the X-ray interaction point. Then, move to the control hutch to use the on-axis microscope for fine adjustment. Move the nozzle tip to the X-ray beam (red X) and write down the absolut nozzle position. From there move the nozzle in +x to your desired distance.
98 +
99 +(% class="wrapped" %)
100 +|=(((
91 91  \\
102 +)))|=(((
103 +direction
104 +)))|=(((
105 +recommended step size
106 +for visual adjustment
107 +)))|=(% colspan="1" %)(((
108 +recommended step size
109 +during data collection
110 +)))
111 +|(((
112 ++x
113 +)))|(((
114 +away from X-ray interaction point
115 +)))|(((
116 +1.0 - 0.5 mm
117 +)))|(% colspan="1" %)(((
118 +0.1 - 0.2 mm
119 +)))
120 +|(((
121 ++y
122 +)))|(((
123 +up
124 +)))|(((
125 +1.0 - 0.5 mm
126 +)))|(% colspan="1" %)(((
127 +0.1 - 0.2 mm
128 +)))
129 +|(((
130 ++z
131 +)))|(((
132 +towards detector tower, away from tape
133 +)))|(((
134 +0.5 - 0.2 mm
135 +)))|(% colspan="1" %)(((
136 +0.1 - 0.2 mm
137 +be careful with -z, tape could be pushed out of focus
138 +)))
92 92  
140 +(% style="color: rgb(255,0,0);" %)**!!! **(% style="color: rgb(255, 0, 0); color: rgb(0, 0, 0)" %)**During data collection make sure in which direction you are moving, do not move nozzle into beam**(% style="color: rgb(255,0,0);" %)** !!!**(%%)**
141 +\\**
142 +)))
143 +** [[image:attach:kamzik_nozzle.PNG||height="400"]]
144 +
93 93  \\
94 94  
147 +\\
148 +
149 +\\
150 +
95 95  Once the hutch is interlocked, set remote control to the TapeDrive laptop by opening a terminal and executing the following command:
96 96  
97 97   xfreerdp +clipboard /size:1920x1080 /u:Cfelusr /v:cfeld-pcx22611