... ... @@ -74,7 +74,7 @@
74 74  **14.4 TapeDrive operation
75 75  **
76 76  
77 -The TapeDrive is controlled through kamzik. To open kamzik, start the server. A terminal will open, shortly after a small window.
77 +The TapeDrive is controlled through kamzik. To open kamzik, start the server. A terminal will open and shortly after a small window.
78 78  Now start the client. Same thing, a terminal will open, shortly after a small window.
79 79  
80 80  Click 'Devices' in the **client session** to open the control window.
... ... @@ -82,16 +82,74 @@
82 82  
83 83  * **Tape:**
84 84  ** Click on the 'main tape' icon
85 -** Change the velocity by typing anything from 0.3-3.0 mm/s
85 +** Change the velocity by typing anything from **0.3-3.0 mm/s**
86 86  ** Common velocity is 1.0-1.5 mm/s
87 +** kamzik displays the used tape in m and %. At 98% the tape will automatically stop since the rolls need to be exchanged at this point.
88 +\\
87 87  ** [[image:attach:kamzik_tape.PNG||height="400"]]
88 88  
89 89  \\
90 90  
93 +* **Sample nozzle:**
94 +** Click an the 'nozzle' icon.
95 +** (((
96 +The nozzle position can be adjusted in xyz by clicking on the arrows up and down. (Double arrow = 10 steps). Step size can be varied, 1mm steps are rather big and should be used carefully.
97 +The absolute position is **not calibrated** and therefore changes after each restart of kamzik.
98 +\\(% style="color: rgb(255,0,0);" %)**!!! **(% style="color: rgb(255, 0, 0); color: rgb(0, 0, 0)" %)**During data collection make sure in which direction you are moving, do not move nozzle into beam**(% style="color: rgb(255,0,0);" %)** !!!**(%%)
99 +\\Adjust the nozzle position by eye until it nearly touches the tape and roughly meets the X-ray interaction point. Then, move to the control hutch to use the on-axis microscope for fine adjustment. Move the nozzle tip to the beam position (red X) and write down the absolut nozzle position. From there move the nozzle in +x to your desired distance.Different distances combined with the tape speed result in several delay times for time-resolved experiments.
100 +
101 +(% class="wrapped" %)
102 +|=(((
91 91  \\
104 +)))|=(((
105 +direction
106 +)))|=(((
107 +recommended step size
108 +for visual adjustment
109 +)))|=(% colspan="1" %)(((
110 +recommended step size
111 +during data collection
112 +)))
113 +|(((
114 ++x
115 +)))|(((
116 +away from X-ray interaction point
117 +)))|(((
118 +1.0 - 0.5 mm
119 +)))|(% colspan="1" %)(((
120 +0.1 - 0.2 mm
121 +)))
122 +|(((
123 ++y
124 +)))|(((
125 +up
126 +)))|(((
127 +1.0 - 0.5 mm
128 +)))|(% colspan="1" %)(((
129 +0.1 - 0.2 mm
130 +)))
131 +|(((
132 ++z
133 +)))|(((
134 +towards detector tower, away from tape
135 +)))|(((
136 +0.5 - 0.2 mm
137 +)))|(% colspan="1" %)(((
138 +0.1 - 0.2 mm
139 +be careful with -z, tape could be pushed out of focus
140 +)))
92 92  
142 +**
143 +\\**
144 +)))
145 +** [[image:attach:kamzik_nozzle.PNG||height="400"]]
146 +
93 93  \\
94 94  
149 +\\
150 +
151 +\\
152 +
95 95  Once the hutch is interlocked, set remote control to the TapeDrive laptop by opening a terminal and executing the following command:
96 96  
97 97   xfreerdp +clipboard /size:1920x1080 /u:Cfelusr /v:cfeld-pcx22611
... ... @@ -104,8 +104,17 @@
104 104  
105 105  **14.5 Tape roll exchange**
106 106  
107 -**[[image:attach:20211201_094111.jpg||height="250"]][[image:attach:20211201_094101.jpg||height="250"]]**
165 +The tape stops automatically at 98% used length, if you need to change the rolls for orther reasons, stop the tape (velocity = 0.0 mm/s).
108 108  
167 +* Remove the (almost) empty feeder roll by pulling it upwards.
168 +* Cut the tape with some cm overlap to glue to the new roll.
169 +* Get a new tape roll (bottom left on the cart). **Polypropylen 15 µm**
170 +* Cut the first cm of the new tape roll, directly before the black sticky part.
171 +* Use this sticky part to attach the end of the old
172 +
173 +**[[image:attach:20211201_094101.jpg||height="250"]][[image:attach:20211201_094111.jpg||height="250"]]
174 +**
175 +
109 109  \\
110 110  
111 111  \\
Confluence.Code.ConfluencePageClass[0]
Id
... ... @@ -1,1 +1,1 @@
1 -235037709
1 +235039398
URL
... ... @@ -1,1 +1,1 @@
1 -https://confluence.desy.de/spaces/11/pages/235037709/14. Serial Crystallography: CFEL TapeDrive
1 +https://confluence.desy.de/spaces/11/pages/235039398/14. Serial Crystallography: CFEL TapeDrive