... ... @@ -1,7 +1,20 @@
1 -[[image:attach:TapeDrive_beschriftet.PNG||height="400"]][[image:attach:TapeDrive scheme.png||height="400"]]
1 +{{layout}}
2 +{{layout-section ac:type="single"}}
3 +{{layout-cell}}
2 2  
3 -**14.1 Revert from Crystallography to TapeDrive (P11 staff/CFEL)**
4 4  
6 +{{toc/}}
7 +
8 +[[image:attach:TapeDrive_beschriftet-tile.jpg||height="400"]]
9 +
10 +\\
11 +
12 +\\
13 +
14 +\\
15 +
16 +=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.1RevertfromCrystallographytoTapeDrive(P11staff/CFEL)"/}}**14.1 Revert from Crystallography to TapeDrive (P11 staff/CFEL)** ===
17 +
5 5  * remove vortex
6 6  * beamstop to out-position?
7 7  * remove goniometer. The airbearing should be stopped (p11/servermotor/eh.1.01)
... ... @@ -50,7 +50,7 @@
50 50  
51 51  \\
52 52  
53 -**14.2 Sample characteristics and preparation**
66 +=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.2Samplecharacteristicsandpreparation"/}}**14.2 Sample characteristics and preparation** ===
54 54  
55 55  suspension of microcrystals, **at least** 200 µL with 20% crystal density(more is always better), size range: 2-30 µm, **tip:** bring some mother liquor for washing, filtering, diluting etc.
56 56  
... ... @@ -62,17 +62,27 @@
62 62  
63 63  \\
64 64  
65 -**14.3 Connecting sample to ElveFlow, sample exchange
66 -**
78 +=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.3ConnectingsampletoElveFlow,sampleexchange"/}}**14.3 Connecting sample to ElveFlow, sample exchange
79 +** ===
67 67  
68 -[[image:attach:20211201_095033.jpg||thumbnail="true" height="250"]][[image:attach:20211201_095205.jpg||height="250"]][[image:attach:20211201_095232.jpg||height="250"]]
81 +After filtration, screw the ElveFlow connector to your Eppendorf or Falcon tube. Make sure that the yellow-ish line goes to the bottom of your tube. You can adjust the length by unscrewing the blue fitting on top and pushing/pulling the yellow-ish line. **Don't forget to tigthen the blue fitting afterwards!**
69 69  
70 -\\
83 +Be aware to which channel and nozzle line your sample is connected.
71 71  
85 + Red → Channel 1, crystal line (nozzle capillary without marks/with red marks)
86 +
87 + Green → Channel 2, substrate line (nozzle capillary is marked in black)
88 +
89 +[[image:attach:20211201_095033.jpg||thumbnail="true" height="250"]]
90 +
91 +After finishing data collection, flush all lines by connecting an Eppi or syringe filled with ddH,,2,,O.
92 +
93 +[[image:attach:20211201_095232.jpg||height="250"]][[image:attach:20211201_095205.jpg||height="250"]]
94 +
72 72  \\
73 73  
74 -**14.4 TapeDrive operation
75 -**
97 +=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.4TapeDriveoperation"/}}**14.4 TapeDrive operation
98 +** ===
76 76  
77 77  The TapeDrive is controlled through kamzik. To open kamzik, start the server. A terminal will open and shortly after a small window.
78 78  Now start the client. Same thing, a terminal will open, shortly after a small window.
... ... @@ -80,24 +80,26 @@
80 80  Click 'Devices' in the **client session** to open the control window.
81 81  [[image:attach:screen.PNG||height="250"]][[image:attach:server client.PNG||height="250"]]
82 82  
83 -* **Tape:**
84 -** Click on the 'main tape' icon
85 -** Change the velocity by typing anything from **0.3-3.0 mm/s**
86 -** Common velocity is 1.0-1.5 mm/s
87 -** kamzik displays the used tape in m and %. At 98% the tape will automatically stop since the rolls need to be exchanged at this point.
88 -\\
89 -** [[image:attach:kamzik_tape.PNG||height="400"]]
106 +**Tape**
90 90  
108 +* Click on the 'main tape' icon
109 +* Change the velocity by typing anything from **0.3-3.0 mm/s**
110 +* Common velocity is 1.0-1.5 mm/s
111 +* kamzik displays the used tape in m and %. At 98% the tape will automatically stop since the rolls need to be exchanged at this point.
91 91  \\
113 +* [[image:attach:kamzik_tape.PNG||height="400"]]
92 92  
93 -* **Sample nozzle:**
94 -** Click an the 'nozzle' icon.
95 -** (((
115 +==== {{id name="14.SerialCrystallography:CFELTapeDrive-Samplenozzle"/}}**Sample nozzle** ====
116 +
117 +* Click an the 'nozzle' icon.
118 +* (((
96 96  The nozzle position can be adjusted in xyz by clicking on the arrows up and down. (Double arrow = 10 steps). Step size can be varied, 1mm steps are rather big and should be used carefully.
97 97  The absolute position is **not calibrated** and therefore changes after each restart of kamzik.
98 -\\(% style="color: rgb(255,0,0);" %)**!!! **(% style="color: rgb(255, 0, 0); color: rgb(0, 0, 0)" %)**During data collection make sure in which direction you are moving, do not move nozzle into beam**(% style="color: rgb(255,0,0);" %)** !!!**(%%)
99 -\\Adjust the nozzle position by eye until it nearly touches the tape and roughly meets the X-ray interaction point. Then, move to the control hutch to use the on-axis microscope for fine adjustment. Move the nozzle tip to the beam position (red X) and write down the absolut nozzle position. From there move the nozzle in +x to your desired distance.Different distances combined with the tape speed result in several delay times for time-resolved experiments.
100 100  
122 +
123 +(% style="color: rgb(255,0,0);" %)**!!! **(% style="color: rgb(255, 0, 0); color: rgb(0, 0, 0)" %)**During data collection make sure in which direction you are moving, do not move nozzle into beam**(% style="color: rgb(255,0,0);" %)** !!!**(%%)
124 +\\Adjust the nozzle position by eye until it nearly touches the tape and roughly meets the X-ray interaction point. Then, move to the control hutch to use the on-axis microscope for fine adjustment. Move the nozzle tip to the beam position (red X) and write down the absolut nozzle position. From there move the nozzle in +x to your desired distance.Different distances combined with the tape speed result in several delay times for time-resolved experiments.
125 +
101 101  (% class="wrapped" %)
102 102  |=(((
103 103  \\
... ... @@ -142,12 +142,20 @@
142 142  **
143 143  \\**
144 144  )))
145 -** [[image:attach:kamzik_nozzle.PNG||height="400"]]
170 +* [[image:attach:kamzik_nozzle.PNG||height="400"]]
146 146  
172 +==== {{id name="14.SerialCrystallography:CFELTapeDrive-Temperaturecontrol/Chiller"/}}**Temperature control/Chiller** ====
173 +
174 +* The chiller circulates fluid through the gas nozzle.
175 +
147 147  \\
148 148  
178 +==== {{id name="14.SerialCrystallography:CFELTapeDrive-Gasnozzle"/}}**Gas nozzle** ====
179 +
149 149  \\
150 150  
182 +==== {{id name="14.SerialCrystallography:CFELTapeDrive-IRcamera"/}}**IR camera** ====
183 +
151 151  \\
152 152  
153 153  Once the hutch is interlocked, set remote control to the TapeDrive laptop by opening a terminal and executing the following command:
... ... @@ -160,17 +160,23 @@
160 160  
161 161  \\
162 162  
163 -**14.5 Tape roll exchange**
196 +=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.5Taperollexchange"/}}**14.5 Tape roll exchange** ===
164 164  
165 165  The tape stops automatically at 98% used length, if you need to change the rolls for orther reasons, stop the tape (velocity = 0.0 mm/s).
166 166  
167 167  * Remove the (almost) empty feeder roll by pulling it upwards.
168 -* Cut the tape with some cm overlap to glue to the new roll.
201 +* Cut the tape with some cm overlap to glue to the new roll. **Do not discard empty roll!**
169 169  * Get a new tape roll (bottom left on the cart). **Polypropylen 15 µm**
170 -* Cut the first cm of the new tape roll, directly before the black sticky part.
171 -* Use this sticky part to attach the end of the old
203 +* Cut the first cm of the new tape roll, directly behind the black sticky part.
204 +* Place new roll on roll holder and use sticky tape to attach the end of the old tape to the beginning of the new roll.
205 +* Open magnetic tape stabilizer on the **left side**. //!!!BEAMSTOP!!!//
206 +* Start the tape (3mm/s). Once the connetion side has arrived at the collector roll, stop the tape (0mm/s).
207 +* Again, cut the tape directly behind the connection side. Replace the the old roll with the **empty **one. Old roll can be discarded.
208 +* Attach new tape to the empty roll by using some sticky tape.
209 +* Close tape stabilizer.
210 +* Reset tape counter in kamzik by clicking "<Call>" in "Main-Tape".
172 172  
173 -**[[image:attach:20211201_094101.jpg||height="250"]][[image:attach:20211201_094111.jpg||height="250"]]
212 +**[[image:attach:20211201_094101.jpg||height="250"]][[image:attach:20211201_094111.jpg||height="250"]][[image:attach:tape reset.PNG||height="250"]]
174 174  **
175 175  
176 176  \\
... ... @@ -177,8 +177,210 @@
177 177  
178 178  \\
179 179  
219 +=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.6Nozzleexchange"/}}**14.6 Nozzle exchange** ===
220 +
221 +=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.7Chopper"/}}**14.7 Chopper** ===
222 +
223 +start chopper gui from terminal:
224 +
225 +(% class="code" %)
226 +(((
227 + cd /gpfs/local/shared/ChopperControl
228 +)))
229 +
230 +(% class="code" %)
231 +(((
232 + python chopperControl.py
233 +)))
234 +
235 +----
236 +{{/layout-cell}}
237 +{{/layout-section}}
238 +
239 +{{layout-section ac:type="two_equal"}}
240 +{{layout-cell}}
241 + [[image:attach:20211212_233140.jpg||height="250"]]
242 +{{/layout-cell}}
243 +
244 +{{layout-cell}}
245 +**Chopper → no Chopper**
246 +
247 +* decrease frequency by max. 10 Hz steps\\
248 +** click arrow down 10x, "Set", arrow down 10x, "Set", ....
249 +* at frequency = 10 Hz, click "Stop"
250 +* set "Hole" to <none> (% style="color: rgb(255,0,0);" %)//Takes some minutes//
251 +
252 +(% style="color: rgb(0,0,0);" %)**no Chopper → Chopper**
253 +
254 +* (% style="color: rgb(0,0,0);" %)set "Hole" to <Inner hole (largest)> (% style="color: rgb(255,0,0);" %)//Takes some minutes//
255 +* (% style="color: rgb(0,0,0);" %)increase frequency by max. 10 Hz steps
256 +** (% style="color: rgb(0,0,0);" %)click arrow up 10x, "Set", arrow up 10x, "Set", .....
257 +* (% style="color: rgb(0,0,0);" %)//usually 133 Hz, results in exposure time of 3.7 ms//
258 +{{/layout-cell}}
259 +{{/layout-section}}
260 +
261 +{{layout-section ac:type="single"}}
262 +{{layout-cell}}
263 +=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.8Datacollection"/}}**14.8 Data collection** ===
264 +
265 +Once the hutch is interlocked, set remote control to the TapeDrive laptop by opening a terminal and executing the following command:
266 +
267 +(% class="code" %)
268 +(((
269 + xfreerdp +clipboard /size:1920x1080 /u:Cfelusr /v:cfeld-pcx22611
270 + password: written on control laptop (changes every 6 months)
271 +)))
272 +
273 + //>>usually the terminal is open, just repeat previous command with arrow up, enter and type in the password//
274 +
180 180  \\
181 181  
277 +Click "Start" icon in kamzik.
278 +
279 +[[image:attach:kamzik_run-tile.jpg||height="400"]]
280 +
281 +----
282 +{{/layout-cell}}
283 +{{/layout-section}}
284 +
285 +{{layout-section ac:type="two_equal"}}
286 +{{layout-cell}}
287 +**Check list:**
288 +
289 +(% style="color: rgb(255,0,0);" %)//always after entering the hutch!//
290 +
291 +* Open BS1  (//CrystalControl)//
292 +* Put collimator in (//CrystalControl)//
293 +* Check beamstop position = 27 000 (//Atk panel p11/motor/eh.3.09)//
294 +* Open fast shutter (//CrystalControl)//
295 +* Check diode voltage ~~ 0.0 (//kamzik Run/Diode voltage)//
296 +* Close fast shutter (//CrystalControl)//
297 +
182 182  \\
183 183  
300 +**Set run parameters and go 🏃💥**
301 +
302 +1. Set your desired No. of frames
303 +1. Match exposure time with chopper setting
304 +1. Check trigger setting and match to chopper setting
305 +1. Set detector distance to 200.0 mm, confirm with //enter// and click "<Call>" button next to it.
306 +1. To start the run, click "<Call>" next to //exposure time//
307 +
184 184  \\
309 +
310 +(% style="color: rgb(153,51,102);" %)**//Current bug~://**
311 +
312 +(% style="color: rgb(0,0,0);" %)You need to "Stop" run in kamzik manually after all images have been acquired. This moves the shield up and closes the fast shutter. Afterwards you can start a new run by clicking "<Call>".
313 +
314 +
315 +(% style="color: rgb(0,0,0);" %)Check progress in terminal with:
316 +
317 +(% class="code" %)
318 +(((
319 + cd /gpfs/current/raw/scan_frames
320 + watch "ls -lrt <run number>/ | tail"
321 +\\\\\\\\
322 +)))
323 +{{/layout-cell}}
324 +
325 +{{layout-cell}}
326 +(% class="wrapped relative-table" style="width: 100.0%;" %)
327 +|=(((
328 +Parameter
329 +)))|=(((
330 +explanation
331 +)))|=(% colspan="1" %)(((
332 +value
333 +)))
334 +|(((
335 +Images number
336 +)))|(((
337 +Set number of frames you want to collect
338 +)))|(% colspan="1" %)(((
339 +usually 100k - 200k, but adjust according to hitrate
340 +
341 +orientation: 10 % hitrate needs at least 100k images
342 +)))
343 +|(((
344 +exposure
345 +)))|(((
346 +exposure time of Eiger
347 +
348 +\\
349 +)))|(% colspan="1" %)(((
350 +chopper at 133 Hz → 3.7 ms
351 +
352 +chopper at 66 Hz   → 7.7 ms
353 +
354 +no chopper            → 7.7 ms
355 +
356 +----
357 +)))
358 +|(((
359 +run id
360 +)))|(((
361 +changes automatically
362 +)))|(% colspan="1" %)(((
363 +\\
364 +)))
365 +|(% colspan="1" %)(((
366 +Detector distance
367 +)))|(% colspan="1" %)(((
368 +current detector postion readout
369 +)))|(% colspan="1" %)(((
370 +\\
371 +)))
372 +|(% colspan="1" %)(((
373 +move_detector (distance)
374 +)))|(% colspan="1" %)(((
375 +Set detector distance
376 +)))|(% colspan="1" %)(((
377 +200.0 mm   (% style="color: rgb(255,0,0);" %)confirm with //enter//, click "<Call>"
378 +)))
379 +|(% colspan="1" %)(((
380 +Diode voltage
381 +)))|(% colspan="1" %)(((
382 +read out of diode
383 +)))|(% colspan="1" %)(((
384 +**~~ 0 !!! higher values indicate missaligned beam stop!!! 
385 +Call local contact
386 +**
387 +)))
388 +{{/layout-cell}}
389 +{{/layout-section}}
390 +
391 +{{layout-section ac:type="single"}}
392 +{{layout-cell}}
393 +----
394 +
395 +=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.9Troubleshooting"/}}14.9 Troubleshooting ===
396 +
397 +* (((
398 +==== {{id name="14.SerialCrystallography:CFELTapeDrive-Tapestopped"/}}Tape stopped ====
399 +
400 +* Tape roll empty?
401 +* motor connected (icon green)?
402 +)))
403 +
404 +\\
405 +
406 +* (((
407 +==== {{id name="14.SerialCrystallography:CFELTapeDrive-Nozzlemotorsarenotresponding"/}}Nozzle motors are not responding ====
408 +
409 +* displayed in green=connected?
410 +* nozzle holder colliding somewhere?
411 +* motor limits reached? → position displayed in red
412 +)))
413 +
414 +\\
415 +
416 +* (((
417 +==== {{id name="14.SerialCrystallography:CFELTapeDrive-Sampleflowisnotrespondingtohigherpressure"/}}Sample flow is not responding to higher pressure ====
418 +
419 +* channel turned on?
420 +* Sample empty?
421 +* Nozzle or lines clogged? → flush with water, change nozzle
422 +)))
423 +{{/layout-cell}}
424 +{{/layout-section}}
425 +{{/layout}}
Confluence.Code.ConfluencePageClass[0]
Id
... ... @@ -1,1 +1,1 @@
1 -235039424
1 +236824589
URL
... ... @@ -1,1 +1,1 @@
1 -https://confluence.desy.de/spaces/11/pages/235039424/14. Serial Crystallography: CFEL TapeDrive
1 +https://confluence.desy.de/spaces/11/pages/236824589/14. Serial Crystallography: CFEL TapeDrive