... ... @@ -79,14 +79,16 @@
79 79  
80 80  Be aware to which channel and nozzle line your sample is connected.
81 81  
82 - Red → Channel 1, crystal line (nozzle capillary without marks)
85 + Red → Channel 1, crystal line (nozzle capillary without marks/with red marks)
83 83  
84 84   Green → Channel 2, substrate line (nozzle capillary is marked in black)
85 85  
86 86  [[image:attach:20211201_095033.jpg||thumbnail="true" height="250"]]
87 87  
88 -\\
91 +After finishing data collection, flush all lines by connecting an Eppi or syringe filled with ddH,,2,,O.
89 89  
93 +[[image:attach:20211201_095232.jpg||height="250"]][[image:attach:20211201_095205.jpg||height="250"]]
94 +
90 90  \\
91 91  
92 92  === {{id name="14.SerialCrystallography:CFELTapeDrive-14.4TapeDriveoperation"/}}**14.4 TapeDrive operation
... ... @@ -113,9 +113,11 @@
113 113  * (((
114 114  The nozzle position can be adjusted in xyz by clicking on the arrows up and down. (Double arrow = 10 steps). Step size can be varied, 1mm steps are rather big and should be used carefully.
115 115  The absolute position is **not calibrated** and therefore changes after each restart of kamzik.
116 -\\(% style="color: rgb(255,0,0);" %)**!!! **(% style="color: rgb(255, 0, 0); color: rgb(0, 0, 0)" %)**During data collection make sure in which direction you are moving, do not move nozzle into beam**(% style="color: rgb(255,0,0);" %)** !!!**(%%)
117 -\\Adjust the nozzle position by eye until it nearly touches the tape and roughly meets the X-ray interaction point. Then, move to the control hutch to use the on-axis microscope for fine adjustment. Move the nozzle tip to the beam position (red X) and write down the absolut nozzle position. From there move the nozzle in +x to your desired distance.Different distances combined with the tape speed result in several delay times for time-resolved experiments.
118 118  
122 +
123 +(% style="color: rgb(255,0,0);" %)**!!! **(% style="color: rgb(255, 0, 0); color: rgb(0, 0, 0)" %)**During data collection make sure in which direction you are moving, do not move nozzle into beam**(% style="color: rgb(255,0,0);" %)** !!!**(%%)
124 +\\Adjust the nozzle position by eye until it nearly touches the tape and roughly meets the X-ray interaction point. Then, move to the control hutch to use the on-axis microscope for fine adjustment. Move the nozzle tip to the beam position (red X) and write down the absolut nozzle position. From there move the nozzle in +x to your desired distance.Different distances combined with the tape speed result in several delay times for time-resolved experiments.
125 +
119 119  (% class="wrapped" %)
120 120  |=(((
121 121  \\
... ... @@ -164,12 +164,16 @@
164 164  
165 165  ==== {{id name="14.SerialCrystallography:CFELTapeDrive-Temperaturecontrol/Chiller"/}}**Temperature control/Chiller** ====
166 166  
167 -\\
174 +* The chiller circulates fluid through the gas nozzle.
168 168  
169 169  \\
170 170  
178 +==== {{id name="14.SerialCrystallography:CFELTapeDrive-Gasnozzle"/}}**Gas nozzle** ====
179 +
171 171  \\
172 172  
182 +==== {{id name="14.SerialCrystallography:CFELTapeDrive-IRcamera"/}}**IR camera** ====
183 +
173 173  \\
174 174  
175 175  Once the hutch is interlocked, set remote control to the TapeDrive laptop by opening a terminal and executing the following command:
... ... @@ -182,7 +182,7 @@
182 182  
183 183  \\
184 184  
185 -==== {{id name="14.SerialCrystallography:CFELTapeDrive-14.5Taperollexchange"/}}**14.5 Tape roll exchange** ====
196 +=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.5Taperollexchange"/}}**14.5 Tape roll exchange** ===
186 186  
187 187  The tape stops automatically at 98% used length, if you need to change the rolls for orther reasons, stop the tape (velocity = 0.0 mm/s).
188 188  
... ... @@ -189,15 +189,16 @@
189 189  * Remove the (almost) empty feeder roll by pulling it upwards.
190 190  * Cut the tape with some cm overlap to glue to the new roll. **Do not discard empty roll!**
191 191  * Get a new tape roll (bottom left on the cart). **Polypropylen 15 µm**
192 -* Cut the first cm of the new tape roll, directly before the black sticky part.
193 -* Use this sticky part to attach the end of the old tape to the new roll. Place new roll on roll holder.
194 -* Open magnetic tape stabilizer.
195 -* Start the tape (3mm/s). Once the connetion side (black sticky part) has arrived at the collector roll, stop the tape (0mm/s).
196 -* Again, cut the tape directly in front of the black mark. Replace the the old roll with the **empty **one. Old roll can be discarded.
197 -* Attach new tape to the empty roll by using the black sticky part.
203 +* Cut the first cm of the new tape roll, directly behind the black sticky part.
204 +* Place new roll on roll holder and use sticky tape to attach the end of the old tape to the beginning of the new roll.
205 +* Open magnetic tape stabilizer on the **left side**. //!!!BEAMSTOP!!!//
206 +* Start the tape (3mm/s). Once the connetion side has arrived at the collector roll, stop the tape (0mm/s).
207 +* Again, cut the tape directly behind the connection side. Replace the the old roll with the **empty **one. Old roll can be discarded.
208 +* Attach new tape to the empty roll by using some sticky tape.
198 198  * Close tape stabilizer.
210 +* Reset tape counter in kamzik by clicking "<Call>" in "Main-Tape".
199 199  
200 -**[[image:attach:20211201_094101.jpg||height="250"]][[image:attach:20211201_094111.jpg||height="250"]]
212 +**[[image:attach:20211201_094101.jpg||height="250"]][[image:attach:20211201_094111.jpg||height="250"]][[image:attach:tape reset.PNG||height="250"]]
201 201  **
202 202  
203 203  \\
... ... @@ -204,8 +204,131 @@
204 204  
205 205  \\
206 206  
219 +=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.6Nozzleexchange"/}}**14.6 Nozzle exchange** ===
220 +
207 207  \\
208 208  
223 +=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.7Datacollection"/}}**14.7 Data collection** ===
224 +
225 +Once the hutch is interlocked, set remote control to the TapeDrive laptop by opening a terminal and executing the following command:
226 +
227 + xfreerdp +clipboard /size:1920x1080 /u:Cfelusr /v:cfeld-pcx22611
228 + password: //written on control laptop (changes every 6 months)//
229 +
230 + //>>usually the terminal is open, just repeat previous command with arrow up, enter and type in the password//
231 +
209 209  \\
210 210  
234 +**Check list:**
235 +
236 +* Open BS1  (//CrystalControl)//
237 +* Put collimator in (//CrystalControl)//
238 +* Check beamstop position = 27 000 (//Atk panel p11/motor/eh.3.09)//
239 +* Open fast shutter (//CrystalControl)//
240 +* Check diode voltage ~~ 0.0 (//kamzik Run/Diode voltage)//
241 +* Close fast shutter (//CrystalControl)//
242 +
211 211  \\
244 +
245 +**Start run in kamzik
246 +**
247 +
248 +Click "Start" icon in kamzik.
249 +
250 +[[image:attach:kamzik_run-tile.jpg||height="400"]]
251 +
252 +----
253 +{{/layout-cell}}
254 +{{/layout-section}}
255 +
256 +{{layout-section ac:type="two_equal"}}
257 +{{layout-cell}}
258 +\\
259 +
260 +1. Set your desired No. of frames
261 +1. Match exposure time with chopper setting
262 +1. Check trigger setting and match to chopper setting
263 +1. Set detector distance to 200.0 mm, confirm with //enter// and click "<Call>" button next to it.
264 +1. To start the run, click "<Call>" next to //exposure time//
265 +{{/layout-cell}}
266 +
267 +{{layout-cell}}
268 +(% class="wrapped" %)
269 +|=(((
270 +Parameter
271 +)))|=(((
272 +explanation
273 +)))|=(% colspan="1" %)(((
274 +value
275 +)))
276 +|(((
277 +Images number
278 +)))|(((
279 +Set number of frames you want to collect
280 +)))|(% colspan="1" %)(((
281 +usually 100k - 200k, but adjust according to hitrate
282 +
283 +orientation: 10 % hitrate needs at least 100k images
284 +)))
285 +|(((
286 +exposure
287 +)))|(((
288 +exposure time of Eiger
289 +
290 +\\
291 +)))|(% colspan="1" %)(((
292 +chopper at 133 Hz → 3.7 ms
293 +
294 +chopper at 66 Hz   → 7.7 ms
295 +
296 +no chopper            → 7.7 ms
297 +
298 +----
299 +)))
300 +|(((
301 +run id
302 +)))|(((
303 +changes automatically
304 +)))|(% colspan="1" %)(((
305 +\\
306 +)))
307 +|(% colspan="1" %)(((
308 +Detector distance
309 +)))|(% colspan="1" %)(((
310 +current detector postion readout
311 +)))|(% colspan="1" %)(((
312 +\\
313 +)))
314 +|(% colspan="1" %)(((
315 +move_detector (distance)
316 +)))|(% colspan="1" %)(((
317 +Set detector distance
318 +)))|(% colspan="1" %)(((
319 +200.0 mm   (% style="color: rgb(255,0,0);" %)confirm with //enter//, click "<Call>"
320 +)))
321 +|(% colspan="1" %)(((
322 +Diode voltage
323 +)))|(% colspan="1" %)(((
324 +read out of diode
325 +)))|(% colspan="1" %)(((
326 +**~~ 0 !!! higher values indicate missaligned beam stop!!!**
327 +)))
328 +{{/layout-cell}}
329 +{{/layout-section}}
330 +
331 +{{layout-section ac:type="single"}}
332 +{{layout-cell}}
333 +=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.8Troubleshooting"/}}14.8 Troubleshooting ===
334 +
335 +* (((
336 +==== {{id name="14.SerialCrystallography:CFELTapeDrive-Tapestopped"/}}Tape stopped ====
337 +)))
338 +* (((
339 +==== {{id name="14.SerialCrystallography:CFELTapeDrive-Nozzlemotorsarenotresponding"/}}Nozzle motors are not responding ====
340 +)))
341 +* (((
342 +==== {{id name="14.SerialCrystallography:CFELTapeDrive-Sampleflowisnotrespondingtohigherpressure"/}}Sample flow is not responding to higher pressure ====
343 +)))
344 +{{/layout-cell}}
345 +{{/layout-section}}
346 +{{/layout}}