Last modified by chatzies on 2025-11-27 11:35

From version 35.1
edited by ahenkel
on 2021-12-12 22:09
Change comment: There is no comment for this version
To version 44.1
edited by ahenkel
on 2021-12-12 23:50
Change comment: There is no comment for this version

Summary

Details

Page properties
Content
... ... @@ -1,9 +1,8 @@
1 +{{layout}}
2 +{{layout-section ac:type="single"}}
3 +{{layout-cell}}
4 +[[image:attach:TapeDrive_beschriftet-tile.jpg||height="250"]]
1 1  
2 -
3 -{{toc/}}
4 -
5 -[[image:attach:TapeDrive_beschriftet-tile.jpg||height="400"]]
6 -
7 7  \\
8 8  
9 9  \\
... ... @@ -115,9 +115,11 @@
115 115  * (((
116 116  The nozzle position can be adjusted in xyz by clicking on the arrows up and down. (Double arrow = 10 steps). Step size can be varied, 1mm steps are rather big and should be used carefully.
117 117  The absolute position is **not calibrated** and therefore changes after each restart of kamzik.
118 -\\(% style="color: rgb(255,0,0);" %)**!!! **(% style="color: rgb(255, 0, 0); color: rgb(0, 0, 0)" %)**During data collection make sure in which direction you are moving, do not move nozzle into beam**(% style="color: rgb(255,0,0);" %)** !!!**(%%)
119 -\\Adjust the nozzle position by eye until it nearly touches the tape and roughly meets the X-ray interaction point. Then, move to the control hutch to use the on-axis microscope for fine adjustment. Move the nozzle tip to the beam position (red X) and write down the absolut nozzle position. From there move the nozzle in +x to your desired distance.Different distances combined with the tape speed result in several delay times for time-resolved experiments.
120 120  
118 +
119 +(% style="color: rgb(255,0,0);" %)**!!! **(% style="color: rgb(255, 0, 0); color: rgb(0, 0, 0)" %)**During data collection make sure in which direction you are moving, do not move nozzle into beam**(% style="color: rgb(255,0,0);" %)** !!!**(%%)
120 +\\Adjust the nozzle position by eye until it nearly touches the tape and roughly meets the X-ray interaction point. Then, move to the control hutch to use the on-axis microscope for fine adjustment. Move the nozzle tip to the beam position (red X) and write down the absolut nozzle position. From there move the nozzle in +x to your desired distance.Different distances combined with the tape speed result in several delay times for time-resolved experiments.
121 +
121 121  (% class="wrapped" %)
122 122  |=(((
123 123  \\
... ... @@ -204,7 +204,7 @@
204 204  * Close tape stabilizer.
205 205  * Reset tape counter in kamzik by clicking "<Call>" in "Main-Tape".
206 206  
207 -**[[image:attach:20211201_094101.jpg||height="250"]][[image:attach:20211201_094111.jpg||height="250"]]
208 +**[[image:attach:20211201_094101.jpg||height="250"]][[image:attach:20211201_094111.jpg||height="250"]][[image:attach:tape reset.PNG||height="250"]]
208 208  **
209 209  
210 210  \\
... ... @@ -211,26 +211,213 @@
211 211  
212 212  \\
213 213  
214 -=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.6Nozzleexchange"/}}14.6 Nozzle exchange ===
215 +=== {{id name="14.SerialCrystallography:CFELTapeDrive-"/}}**
216 +** ===
217 +{{/layout-cell}}
218 +{{/layout-section}}
215 215  
220 +{{layout-section ac:type="single"}}
221 +{{layout-cell}}
222 +=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.7Chopper"/}}**14.7 Chopper** ===
223 +
224 +start chopper gui from terminal:
225 +
226 +(% class="code" %)
227 +(((
228 + cd /gpfs/local/shared/ChopperControl
229 +)))
230 +
231 +(% class="code" %)
232 +(((
233 + python chopperControl.py
234 +)))
235 +{{/layout-cell}}
236 +{{/layout-section}}
237 +
238 +{{layout-section ac:type="two_equal"}}
239 +{{layout-cell}}
240 + [[image:attach:20211212_233140.jpg||height="250"]]
241 +{{/layout-cell}}
242 +
243 +{{layout-cell}}
244 +**Chopper → no Chopper**
245 +
246 +* decrease frequency by max. 10 Hz steps\\
247 +** click arrow down 10x, "Set", arrow down 10x, "Set", ....
248 +* at frequency = 10 Hz, click "Stop"
249 +* set "Hole" to <none> (% style="color: rgb(255,0,0);" %)//Takes some minutes//
250 +
251 +(% style="color: rgb(0,0,0);" %)**no Chopper → Chopper**
252 +
253 +* (% style="color: rgb(0,0,0);" %)set "Hole" to <Inner hole (largest)> (% style="color: rgb(255,0,0);" %)//Takes some minutes//
254 +* (% style="color: rgb(0,0,0);" %)increase frequency by max. 10 Hz steps
255 +** (% style="color: rgb(0,0,0);" %)click arrow up 10x, "Set", arrow up 10x, "Set", .....
256 +* (% style="color: rgb(0,0,0);" %)//usually 133 Hz, results in exposure time of 3.7 ms//
257 +{{/layout-cell}}
258 +{{/layout-section}}
259 +
260 +{{layout-section ac:type="single"}}
261 +{{layout-cell}}
262 +=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.8Datacollection"/}}**14.8 Data collection** ===
263 +
264 +Once the hutch is interlocked, set remote control to the TapeDrive laptop by opening a terminal and executing the following command:
265 +
266 +(% class="code" %)
267 +(((
268 + xfreerdp +clipboard /size:1920x1080 /u:Cfelusr /v:cfeld-pcx22611
269 + password: written on control laptop (changes every 6 months)
270 +)))
271 +
272 + //>>usually the terminal is open, just repeat previous command with arrow up, enter and type in the password//
273 +
216 216  \\
217 217  
218 -=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.7Datacollection"/}}14.7 Data collection ===
276 +Click "Start" icon in kamzik.
219 219  
278 +[[image:attach:kamzik_run-tile.jpg||height="400"]]
279 +
280 +----
281 +{{/layout-cell}}
282 +{{/layout-section}}
283 +
284 +{{layout-section ac:type="two_equal"}}
285 +{{layout-cell}}
286 +**Check list:**
287 +
288 +(% style="color: rgb(255,0,0);" %)//always after entering the hutch!//
289 +
290 +* Open BS1  (//CrystalControl)//
291 +* Put collimator in (//CrystalControl)//
292 +* Check beamstop position = 27 000 (//Atk panel p11/motor/eh.3.09)//
293 +* Open fast shutter (//CrystalControl)//
294 +* Check diode voltage ~~ 0.0 (//kamzik Run/Diode voltage)//
295 +* Close fast shutter (//CrystalControl)//
296 +
220 220  \\
221 221  
222 -=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.8Troubleshooting"/}}14.8 Troubleshooting ===
299 +**Set run parameters and go 🏃💥**
223 223  
301 +1. Set your desired No. of frames
302 +1. Match exposure time with chopper setting
303 +1. Check trigger setting and match to chopper setting
304 +1. Set detector distance to 200.0 mm, confirm with //enter// and click "<Call>" button next to it.
305 +1. To start the run, click "<Call>" next to //exposure time//
306 +
307 +\\
308 +
309 +(% style="color: rgb(153,51,102);" %)**//Current bug~://**
310 +
311 +(% style="color: rgb(0,0,0);" %)You need to "Stop" run in kamzik manually after all images have been acquired. This moves the shield up and closes the fast shutter. Afterwards you can start a new run by clicking "<Call>".
312 +
313 +
314 +(% style="color: rgb(0,0,0);" %)Check progress in terminal with:
315 +
316 +(% class="code" %)
317 +(((
318 + cd /gpfs/current/raw/scan_frames
319 + watch "ls -lrt <run number>/ | tail"
320 +\\\\\\\\
321 +)))
322 +{{/layout-cell}}
323 +
324 +{{layout-cell}}
325 +(% class="wrapped relative-table" style="width: 100.0%;" %)
326 +|=(((
327 +Parameter
328 +)))|=(((
329 +explanation
330 +)))|=(% colspan="1" %)(((
331 +value
332 +)))
333 +|(((
334 +Images number
335 +)))|(((
336 +Set number of frames you want to collect
337 +)))|(% colspan="1" %)(((
338 +usually 100k - 200k, but adjust according to hitrate
339 +
340 +orientation: 10 % hitrate needs at least 100k images
341 +)))
342 +|(((
343 +exposure
344 +)))|(((
345 +exposure time of Eiger
346 +
347 +\\
348 +)))|(% colspan="1" %)(((
349 +chopper at 133 Hz → 3.7 ms
350 +
351 +chopper at 66 Hz   → 7.7 ms
352 +
353 +no chopper            → 7.7 ms
354 +
355 +----
356 +)))
357 +|(((
358 +run id
359 +)))|(((
360 +changes automatically
361 +)))|(% colspan="1" %)(((
362 +\\
363 +)))
364 +|(% colspan="1" %)(((
365 +Detector distance
366 +)))|(% colspan="1" %)(((
367 +current detector postion readout
368 +)))|(% colspan="1" %)(((
369 +\\
370 +)))
371 +|(% colspan="1" %)(((
372 +move_detector (distance)
373 +)))|(% colspan="1" %)(((
374 +Set detector distance
375 +)))|(% colspan="1" %)(((
376 +200.0 mm   (% style="color: rgb(255,0,0);" %)confirm with //enter//, click "<Call>"
377 +)))
378 +|(% colspan="1" %)(((
379 +Diode voltage
380 +)))|(% colspan="1" %)(((
381 +read out of diode
382 +)))|(% colspan="1" %)(((
383 +**~~ 0 !!! higher values indicate missaligned beam stop!!! 
384 +Call local contact
385 +**
386 +)))
387 +{{/layout-cell}}
388 +{{/layout-section}}
389 +
390 +{{layout-section ac:type="single"}}
391 +{{layout-cell}}
392 +----
393 +
394 +=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.9Troubleshooting"/}}14.9 Troubleshooting ===
395 +
224 224  * (((
225 225  ==== {{id name="14.SerialCrystallography:CFELTapeDrive-Tapestopped"/}}Tape stopped ====
398 +
399 +* Tape roll empty?
400 +* motor connected (icon green)?
226 226  )))
402 +
403 +\\
404 +
227 227  * (((
228 228  ==== {{id name="14.SerialCrystallography:CFELTapeDrive-Nozzlemotorsarenotresponding"/}}Nozzle motors are not responding ====
407 +
408 +* displayed in green=connected?
409 +* nozzle holder colliding somewhere?
410 +* motor limits reached? → position displayed in red
229 229  )))
230 -* (((
231 -==== {{id name="14.SerialCrystallography:CFELTapeDrive-Sampleflowisnotrespondingtohigherpressure"/}}Sample flow is not responding to higher pressure ====
232 -)))
233 233  
234 234  \\
235 235  
236 -\\
415 +* (((
416 +==== {{id name="14.SerialCrystallography:CFELTapeDrive-Sampleflowisnotrespondingtohigherpressure"/}}Sample flow is not responding to higher pressure ====
417 +
418 +* channel turned on?
419 +* Sample empty?
420 +* Nozzle or lines clogged? → flush with water, change nozzle
421 +)))
422 +{{/layout-cell}}
423 +{{/layout-section}}
424 +{{/layout}}
Confluence.Code.ConfluencePageClass[0]
Id
... ... @@ -1,1 +1,1 @@
1 -236824561
1 +236824593
URL
... ... @@ -1,1 +1,1 @@
1 -https://confluence.desy.de/spaces/11/pages/236824561/14. Serial Crystallography: CFEL TapeDrive
1 +https://confluence.desy.de/spaces/11/pages/236824593/14. Serial Crystallography: CFEL TapeDrive