Last modified by chatzies on 2025-11-27 11:35

From version 36.1
edited by ahenkel
on 2021-12-12 22:53
Change comment: There is no comment for this version
To version 43.1
edited by ahenkel
on 2021-12-12 23:49
Change comment: There is no comment for this version

Summary

Details

Page properties
Content
... ... @@ -1,7 +1,6 @@
1 -
2 -
3 -{{toc/}}
4 -
1 +{{layout}}
2 +{{layout-section ac:type="single"}}
3 +{{layout-cell}}
5 5  [[image:attach:TapeDrive_beschriftet-tile.jpg||height="400"]]
6 6  
7 7  \\
... ... @@ -120,7 +120,6 @@
120 120  (% style="color: rgb(255,0,0);" %)**!!! **(% style="color: rgb(255, 0, 0); color: rgb(0, 0, 0)" %)**During data collection make sure in which direction you are moving, do not move nozzle into beam**(% style="color: rgb(255,0,0);" %)** !!!**(%%)
121 121  \\Adjust the nozzle position by eye until it nearly touches the tape and roughly meets the X-ray interaction point. Then, move to the control hutch to use the on-axis microscope for fine adjustment. Move the nozzle tip to the beam position (red X) and write down the absolut nozzle position. From there move the nozzle in +x to your desired distance.Different distances combined with the tape speed result in several delay times for time-resolved experiments.
122 122  
123 -(% class="wrapped" %)
124 124  |=(((
125 125  \\
126 126  )))|=(((
... ... @@ -214,22 +214,79 @@
214 214  \\
215 215  
216 216  === {{id name="14.SerialCrystallography:CFELTapeDrive-14.6Nozzleexchange"/}}**14.6 Nozzle exchange** ===
215 +{{/layout-cell}}
216 +{{/layout-section}}
217 217  
218 -\\
218 +{{layout-section ac:type="single"}}
219 +{{layout-cell}}
220 +=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.7Chopper"/}}**14.7 Chopper** ===
219 219  
220 -=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.7Datacollection"/}}**14.7 Data collection** ===
222 +start chopper gui from terminal:
221 221  
224 +(% class="code" %)
225 +(((
226 + cd /gpfs/local/shared/ChopperControl
227 +)))
228 +
229 +(% class="code" %)
230 +(((
231 + python chopperControl.py
232 +)))
233 +{{/layout-cell}}
234 +{{/layout-section}}
235 +
236 +{{layout-section ac:type="two_equal"}}
237 +{{layout-cell}}
238 + [[image:attach:20211212_233140.jpg||height="250"]]
239 +{{/layout-cell}}
240 +
241 +{{layout-cell}}
242 +**Chopper → no Chopper**
243 +
244 +* decrease frequency by max. 10 Hz steps\\
245 +** click arrow down 10x, "Set", arrow down 10x, "Set", ....
246 +* at frequency = 10 Hz, click "Stop"
247 +* set "Hole" to <none> (% style="color: rgb(255,0,0);" %)//Takes some minutes//
248 +
249 +(% style="color: rgb(0,0,0);" %)**no Chopper → Chopper**
250 +
251 +* (% style="color: rgb(0,0,0);" %)set "Hole" to <Inner hole (largest)> (% style="color: rgb(255,0,0);" %)//Takes some minutes//
252 +* (% style="color: rgb(0,0,0);" %)increase frequency by max. 10 Hz steps
253 +** (% style="color: rgb(0,0,0);" %)click arrow up 10x, "Set", arrow up 10x, "Set", .....
254 +* (% style="color: rgb(0,0,0);" %)//usually 133 Hz, results in exposure time of 3.7 ms//
255 +{{/layout-cell}}
256 +{{/layout-section}}
257 +
258 +{{layout-section ac:type="single"}}
259 +{{layout-cell}}
260 +=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.8Datacollection"/}}**14.8 Data collection** ===
261 +
222 222  Once the hutch is interlocked, set remote control to the TapeDrive laptop by opening a terminal and executing the following command:
223 223  
224 - xfreerdp +clipboard /size:1920x1080 /u:Cfelusr /v:cfeld-pcx22611
225 - password: //written on control laptop (changes every 6 months)//
264 +(% class="code" %)
265 +(((
266 + xfreerdp +clipboard /size:1920x1080 /u:Cfelusr /v:cfeld-pcx22611
267 + password: written on control laptop (changes every 6 months)
268 +)))
226 226  
227 227   //>>usually the terminal is open, just repeat previous command with arrow up, enter and type in the password//
228 228  
229 229  \\
230 230  
274 +Click "Start" icon in kamzik.
275 +
276 +[[image:attach:kamzik_run-tile.jpg||height="400"]]
277 +
278 +----
279 +{{/layout-cell}}
280 +{{/layout-section}}
281 +
282 +{{layout-section ac:type="two_equal"}}
283 +{{layout-cell}}
231 231  **Check list:**
232 232  
286 +(% style="color: rgb(255,0,0);" %)//always after entering the hutch!//
287 +
233 233  * Open BS1  (//CrystalControl)//
234 234  * Put collimator in (//CrystalControl)//
235 235  * Check beamstop position = 27 000 (//Atk panel p11/motor/eh.3.09)//
... ... @@ -239,13 +239,33 @@
239 239  
240 240  \\
241 241  
242 -**Start run in kamzik
243 -**
297 +**Set run parameters and go 🏃💥**
244 244  
245 -Click "Start" icon in kamzik.
299 +1. Set your desired No. of frames
300 +1. Match exposure time with chopper setting
301 +1. Check trigger setting and match to chopper setting
302 +1. Set detector distance to 200.0 mm, confirm with //enter// and click "<Call>" button next to it.
303 +1. To start the run, click "<Call>" next to //exposure time//
246 246  
247 -[[image:attach:kamzik_run-tile.jpg||height="400"]]
305 +\\
248 248  
307 +(% style="color: rgb(153,51,102);" %)**//Current bug~://**
308 +
309 +(% style="color: rgb(0,0,0);" %)You need to "Stop" run in kamzik manually after all images have been acquired. This moves the shield up and closes the fast shutter. Afterwards you can start a new run by clicking "<Call>".
310 +
311 +
312 +(% style="color: rgb(0,0,0);" %)Check progress in terminal with:
313 +
314 +(% class="code" %)
315 +(((
316 + cd /gpfs/current/raw/scan_frames
317 + watch "ls -lrt <run number>/ | tail"
318 +\\\\\\\\
319 +)))
320 +{{/layout-cell}}
321 +
322 +{{layout-cell}}
323 +(% class="wrapped relative-table" style="width: 100.0%;" %)
249 249  |=(((
250 250  Parameter
251 251  )))|=(((
... ... @@ -303,52 +303,45 @@
303 303  )))|(% colspan="1" %)(((
304 304  read out of diode
305 305  )))|(% colspan="1" %)(((
306 -**~~ 0 !!! higher values indicate missaligned beam stop!!!**
381 +**~~ 0 !!! higher values indicate missaligned beam stop!!! 
382 +Call local contact
383 +**
307 307  )))
385 +{{/layout-cell}}
386 +{{/layout-section}}
308 308  
388 +{{layout-section ac:type="single"}}
389 +{{layout-cell}}
309 309  ----
310 310  
311 -(% class="task-list" %)
312 -(((
313 -{{task reference="/Tasks/Task_6" status="InProgress"}}
314 -Set your desired No. of frames
315 -{{/task}}
392 +=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.9Troubleshooting"/}}14.9 Troubleshooting ===
316 316  
317 -{{task reference="/Tasks/Task_7" status="InProgress"}}
318 -Match exposure time with chopper setting
319 -{{/task}}
394 +* (((
395 +==== {{id name="14.SerialCrystallography:CFELTapeDrive-Tapestopped"/}}Tape stopped ====
320 320  
321 -{{task reference="/Tasks/Task_8" status="InProgress"}}
322 -Check trigger setting and match to chopper setting
323 -{{/task}}
324 -
325 -{{task reference="/Tasks/Task_9" status="InProgress"}}
326 -Set detector distance to 200.0 mm, confirm with //enter// and click "<Call>" button next to it.
327 -{{/task}}
328 -
329 -{{task reference="/Tasks/Task_10" status="InProgress"}}
330 -To start the run, click "<Call>" next to //exposure time//
331 -{{/task}}
397 +* Tape roll empty?
398 +* motor connected (icon green)?
332 332  )))
333 333  
334 -----
335 -
336 336  \\
337 337  
338 -\\
339 -
340 -=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.8Troubleshooting"/}}14.8 Troubleshooting ===
341 -
342 342  * (((
343 -==== {{id name="14.SerialCrystallography:CFELTapeDrive-Tapestopped"/}}Tape stopped ====
344 -)))
345 -* (((
346 346  ==== {{id name="14.SerialCrystallography:CFELTapeDrive-Nozzlemotorsarenotresponding"/}}Nozzle motors are not responding ====
405 +
406 +* displayed in green=connected?
407 +* nozzle holder colliding somewhere?
408 +* motor limits reached? → position displayed in red
347 347  )))
348 -* (((
349 -==== {{id name="14.SerialCrystallography:CFELTapeDrive-Sampleflowisnotrespondingtohigherpressure"/}}Sample flow is not responding to higher pressure ====
350 -)))
351 351  
352 352  \\
353 353  
354 -\\
413 +* (((
414 +==== {{id name="14.SerialCrystallography:CFELTapeDrive-Sampleflowisnotrespondingtohigherpressure"/}}Sample flow is not responding to higher pressure ====
415 +
416 +* channel turned on?
417 +* Sample empty?
418 +* Nozzle or lines clogged? → flush with water, change nozzle
419 +)))
420 +{{/layout-cell}}
421 +{{/layout-section}}
422 +{{/layout}}
Confluence.Code.ConfluencePageClass[0]
Id
... ... @@ -1,1 +1,1 @@
1 -236824562
1 +236824590
URL
... ... @@ -1,1 +1,1 @@
1 -https://confluence.desy.de/spaces/11/pages/236824562/14. Serial Crystallography: CFEL TapeDrive
1 +https://confluence.desy.de/spaces/11/pages/236824590/14. Serial Crystallography: CFEL TapeDrive