... ... @@ -1,16 +1,7 @@
1 -{{layout}}
2 -{{layout-section ac:type="single"}}
3 -{{layout-cell}}
4 -[[image:attach:TapeDrive_beschriftet-tile.jpg||height="400"]]
1 +[[image:attach:TapeDrive_beschriftet.PNG||height="400"]][[image:attach:TapeDrive scheme.png||height="400"]]
5 5  
6 -\\
3 +**14.1 Revert from Crystallography to TapeDrive (P11 staff/CFEL)**
7 7  
8 -\\
9 -
10 -\\
11 -
12 -=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.1RevertfromCrystallographytoTapeDrive(P11staff/CFEL)"/}}**14.1 Revert from Crystallography to TapeDrive (P11 staff/CFEL)** ===
13 -
14 14  * remove vortex
15 15  * beamstop to out-position?
16 16  * remove goniometer. The airbearing should be stopped (p11/servermotor/eh.1.01)
... ... @@ -59,7 +59,7 @@
59 59  
60 60  \\
61 61  
62 -=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.2Samplecharacteristicsandpreparation"/}}**14.2 Sample characteristics and preparation** ===
53 +**14.2 Sample characteristics and preparation**
63 63  
64 64  suspension of microcrystals, **at least** 200 µL with 20% crystal density(more is always better), size range: 2-30 µm, **tip:** bring some mother liquor for washing, filtering, diluting etc.
65 65  
... ... @@ -71,113 +71,36 @@
71 71  
72 72  \\
73 73  
74 -=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.3ConnectingsampletoElveFlow,sampleexchange"/}}**14.3 Connecting sample to ElveFlow, sample exchange
75 -** ===
65 +**14.3 Connecting sample to ElveFlow, sample exchange
66 +**
76 76  
77 -After filtration, screw the ElveFlow connector to your Eppendorf or Falcon tube. Make sure that the yellow-ish line goes to the bottom of your tube. You can adjust the length by unscrewing the blue fitting on top and pushing/pulling the yellow-ish line. **Don't forget to tigthen the blue fitting afterwards!**
68 +[[image:attach:20211201_095033.jpg||thumbnail="true" height="250"]][[image:attach:20211201_095205.jpg||height="250"]][[image:attach:20211201_095232.jpg||height="250"]]
78 78  
79 -Be aware to which channel and nozzle line your sample is connected.
70 +\\
80 80  
81 - Red → Channel 1, crystal line (nozzle capillary without marks/with red marks)
82 -
83 - Green → Channel 2, substrate line (nozzle capillary is marked in black)
84 -
85 -[[image:attach:20211201_095033.jpg||thumbnail="true" height="250"]]
86 -
87 -After finishing data collection, flush all lines by connecting an Eppi or syringe filled with ddH,,2,,O.
88 -
89 -[[image:attach:20211201_095232.jpg||height="250"]][[image:attach:20211201_095205.jpg||height="250"]]
90 -
91 91  \\
92 92  
93 -=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.4TapeDriveoperation"/}}**14.4 TapeDrive operation
94 -** ===
74 +**14.4 TapeDrive operation
75 +**
95 95  
96 -The TapeDrive is controlled through kamzik. To open kamzik, start the server. A terminal will open and shortly after a small window.
77 +The TapeDrive is controlled through kamzik. To open kamzik, start the server. A terminal will open, shortly after a small window.
97 97  Now start the client. Same thing, a terminal will open, shortly after a small window.
98 98  
99 99  Click 'Devices' in the **client session** to open the control window.
100 100  [[image:attach:screen.PNG||height="250"]][[image:attach:server client.PNG||height="250"]]
101 101  
102 -**Tape**
83 +* **Tape:**
84 +** Click on the 'main tape' icon
85 +** Change the velocity by typing anything from 0.3-3.0 mm/s
86 +** Common velocity is 1.0-1.5 mm/s
87 +** [[image:attach:kamzik_tape.PNG||height="400"]]
103 103  
104 -* Click on the 'main tape' icon
105 -* Change the velocity by typing anything from **0.3-3.0 mm/s**
106 -* Common velocity is 1.0-1.5 mm/s
107 -* kamzik displays the used tape in m and %. At 98% the tape will automatically stop since the rolls need to be exchanged at this point.
108 108  \\
109 -* [[image:attach:kamzik_tape.PNG||height="400"]]
110 110  
111 -==== {{id name="14.SerialCrystallography:CFELTapeDrive-Samplenozzle"/}}**Sample nozzle** ====
112 -
113 -* Click an the 'nozzle' icon.
114 -* (((
115 -The nozzle position can be adjusted in xyz by clicking on the arrows up and down. (Double arrow = 10 steps). Step size can be varied, 1mm steps are rather big and should be used carefully.
116 -The absolute position is **not calibrated** and therefore changes after each restart of kamzik.
117 -
118 -
119 -(% style="color: rgb(255,0,0);" %)**!!! **(% style="color: rgb(255, 0, 0); color: rgb(0, 0, 0)" %)**During data collection make sure in which direction you are moving, do not move nozzle into beam**(% style="color: rgb(255,0,0);" %)** !!!**(%%)
120 -\\Adjust the nozzle position by eye until it nearly touches the tape and roughly meets the X-ray interaction point. Then, move to the control hutch to use the on-axis microscope for fine adjustment. Move the nozzle tip to the beam position (red X) and write down the absolut nozzle position. From there move the nozzle in +x to your desired distance.Different distances combined with the tape speed result in several delay times for time-resolved experiments.
121 -
122 -|=(((
123 123  \\
124 -)))|=(((
125 -direction
126 -)))|=(((
127 -recommended step size
128 -for visual adjustment
129 -)))|=(% colspan="1" %)(((
130 -recommended step size
131 -during data collection
132 -)))
133 -|(((
134 -+x
135 -)))|(((
136 -away from X-ray interaction point
137 -)))|(((
138 -1.0 - 0.5 mm
139 -)))|(% colspan="1" %)(((
140 -0.1 - 0.2 mm
141 -)))
142 -|(((
143 -+y
144 -)))|(((
145 -up
146 -)))|(((
147 -1.0 - 0.5 mm
148 -)))|(% colspan="1" %)(((
149 -0.1 - 0.2 mm
150 -)))
151 -|(((
152 -+z
153 -)))|(((
154 -towards detector tower, away from tape
155 -)))|(((
156 -0.5 - 0.2 mm
157 -)))|(% colspan="1" %)(((
158 -0.1 - 0.2 mm
159 -be careful with -z, tape could be pushed out of focus
160 -)))
161 161  
162 -**
163 -\\**
164 -)))
165 -* [[image:attach:kamzik_nozzle.PNG||height="400"]]
166 -
167 -==== {{id name="14.SerialCrystallography:CFELTapeDrive-Temperaturecontrol/Chiller"/}}**Temperature control/Chiller** ====
168 -
169 -* The chiller circulates fluid through the gas nozzle.
170 -
171 171  \\
172 172  
173 -==== {{id name="14.SerialCrystallography:CFELTapeDrive-Gasnozzle"/}}**Gas nozzle** ====
174 -
175 -\\
176 -
177 -==== {{id name="14.SerialCrystallography:CFELTapeDrive-IRcamera"/}}**IR camera** ====
178 -
179 -\\
180 -
181 181  Once the hutch is interlocked, set remote control to the TapeDrive laptop by opening a terminal and executing the following command:
182 182  
183 183   xfreerdp +clipboard /size:1920x1080 /u:Cfelusr /v:cfeld-pcx22611
... ... @@ -188,235 +188,16 @@
188 188  
189 189  \\
190 190  
191 -=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.5Taperollexchange"/}}**14.5 Tape roll exchange** ===
105 +**14.5 Tape roll exchange**
192 192  
193 -The tape stops automatically at 98% used length, if you need to change the rolls for orther reasons, stop the tape (velocity = 0.0 mm/s).
107 +**[[image:attach:20211201_094111.jpg||height="250"]][[image:attach:20211201_094101.jpg||height="250"]]**
194 194  
195 -* Remove the (almost) empty feeder roll by pulling it upwards.
196 -* Cut the tape with some cm overlap to glue to the new roll. **Do not discard empty roll!**
197 -* Get a new tape roll (bottom left on the cart). **Polypropylen 15 µm**
198 -* Cut the first cm of the new tape roll, directly behind the black sticky part.
199 -* Place new roll on roll holder and use sticky tape to attach the end of the old tape to the beginning of the new roll.
200 -* Open magnetic tape stabilizer on the **left side**. //!!!BEAMSTOP!!!//
201 -* Start the tape (3mm/s). Once the connetion side has arrived at the collector roll, stop the tape (0mm/s).
202 -* Again, cut the tape directly behind the connection side. Replace the the old roll with the **empty **one. Old roll can be discarded.
203 -* Attach new tape to the empty roll by using some sticky tape.
204 -* Close tape stabilizer.
205 -* Reset tape counter in kamzik by clicking "<Call>" in "Main-Tape".
206 -
207 -**[[image:attach:20211201_094101.jpg||height="250"]][[image:attach:20211201_094111.jpg||height="250"]][[image:attach:tape reset.PNG||height="250"]]
208 -**
209 -
210 210  \\
211 211  
212 212  \\
213 213  
214 -=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.6Nozzleexchange"/}}**14.6 Nozzle exchange** ===
215 -{{/layout-cell}}
216 -{{/layout-section}}
217 -
218 -{{layout-section ac:type="single"}}
219 -{{layout-cell}}
220 -=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.7Chopper"/}}**14.7 Chopper** ===
221 -
222 -start chopper gui from terminal:
223 -
224 -(% class="code" %)
225 -(((
226 - cd /gpfs/local/shared/ChopperControl
227 -)))
228 -
229 -(% class="code" %)
230 -(((
231 - python chopperControl.py
232 -)))
233 -{{/layout-cell}}
234 -{{/layout-section}}
235 -
236 -{{layout-section ac:type="two_equal"}}
237 -{{layout-cell}}
238 - [[image:attach:20211212_233140.jpg||height="250"]]
239 -{{/layout-cell}}
240 -
241 -{{layout-cell}}
242 -**Chopper → no Chopper**
243 -
244 -* decrease frequency by max. 10 Hz steps\\
245 -** click arrow down 10x, "Set", arrow down 10x, "Set", ....
246 -* at frequency = 10 Hz, click "Stop"
247 -* set "Hole" to <none> (% style="color: rgb(255,0,0);" %)//Takes some minutes//
248 -
249 -(% style="color: rgb(0,0,0);" %)**no Chopper → Chopper**
250 -
251 -* (% style="color: rgb(0,0,0);" %)set "Hole" to <Inner hole (largest)> (% style="color: rgb(255,0,0);" %)//Takes some minutes//
252 -* (% style="color: rgb(0,0,0);" %)increase frequency by max. 10 Hz steps
253 -** (% style="color: rgb(0,0,0);" %)click arrow up 10x, "Set", arrow up 10x, "Set", .....
254 -* (% style="color: rgb(0,0,0);" %)//usually 133 Hz, results in exposure time of 3.7 ms//
255 -{{/layout-cell}}
256 -{{/layout-section}}
257 -
258 -{{layout-section ac:type="single"}}
259 -{{layout-cell}}
260 -=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.8Datacollection"/}}**14.8 Data collection** ===
261 -
262 -Once the hutch is interlocked, set remote control to the TapeDrive laptop by opening a terminal and executing the following command:
263 -
264 -(% class="code" %)
265 -(((
266 - xfreerdp +clipboard /size:1920x1080 /u:Cfelusr /v:cfeld-pcx22611
267 - password: written on control laptop (changes every 6 months)
268 -)))
269 -
270 - //>>usually the terminal is open, just repeat previous command with arrow up, enter and type in the password//
271 -
272 272  \\
273 273  
274 -Click "Start" icon in kamzik.
275 -
276 -[[image:attach:kamzik_run-tile.jpg||height="400"]]
277 -
278 -----
279 -{{/layout-cell}}
280 -{{/layout-section}}
281 -
282 -{{layout-section ac:type="two_equal"}}
283 -{{layout-cell}}
284 -**Check list:**
285 -
286 -(% style="color: rgb(255,0,0);" %)//always after entering the hutch!//
287 -
288 -* Open BS1  (//CrystalControl)//
289 -* Put collimator in (//CrystalControl)//
290 -* Check beamstop position = 27 000 (//Atk panel p11/motor/eh.3.09)//
291 -* Open fast shutter (//CrystalControl)//
292 -* Check diode voltage ~~ 0.0 (//kamzik Run/Diode voltage)//
293 -* Close fast shutter (//CrystalControl)//
294 -
295 295  \\
296 296  
297 -**Set run parameters and go 🏃💥**
298 -
299 -1. Set your desired No. of frames
300 -1. Match exposure time with chopper setting
301 -1. Check trigger setting and match to chopper setting
302 -1. Set detector distance to 200.0 mm, confirm with //enter// and click "<Call>" button next to it.
303 -1. To start the run, click "<Call>" next to //exposure time//
304 -
305 305  \\
306 -
307 -(% style="color: rgb(153,51,102);" %)**//Current bug~://**
308 -
309 -(% style="color: rgb(0,0,0);" %)You need to "Stop" run in kamzik manually after all images have been acquired. This moves the shield up and closes the fast shutter. Afterwards you can start a new run by clicking "<Call>".
310 -
311 -
312 -(% style="color: rgb(0,0,0);" %)Check progress in terminal with:
313 -
314 -(% class="code" %)
315 -(((
316 - cd /gpfs/current/raw/scan_frames
317 - watch "ls -lrt <run number>/ | tail"
318 -\\\\\\\\
319 -)))
320 -{{/layout-cell}}
321 -
322 -{{layout-cell}}
323 -(% class="wrapped relative-table" style="width: 100.0%;" %)
324 -|=(((
325 -Parameter
326 -)))|=(((
327 -explanation
328 -)))|=(% colspan="1" %)(((
329 -value
330 -)))
331 -|(((
332 -Images number
333 -)))|(((
334 -Set number of frames you want to collect
335 -)))|(% colspan="1" %)(((
336 -usually 100k - 200k, but adjust according to hitrate
337 -
338 -orientation: 10 % hitrate needs at least 100k images
339 -)))
340 -|(((
341 -exposure
342 -)))|(((
343 -exposure time of Eiger
344 -
345 -\\
346 -)))|(% colspan="1" %)(((
347 -chopper at 133 Hz → 3.7 ms
348 -
349 -chopper at 66 Hz   → 7.7 ms
350 -
351 -no chopper            → 7.7 ms
352 -
353 -----
354 -)))
355 -|(((
356 -run id
357 -)))|(((
358 -changes automatically
359 -)))|(% colspan="1" %)(((
360 -\\
361 -)))
362 -|(% colspan="1" %)(((
363 -Detector distance
364 -)))|(% colspan="1" %)(((
365 -current detector postion readout
366 -)))|(% colspan="1" %)(((
367 -\\
368 -)))
369 -|(% colspan="1" %)(((
370 -move_detector (distance)
371 -)))|(% colspan="1" %)(((
372 -Set detector distance
373 -)))|(% colspan="1" %)(((
374 -200.0 mm   (% style="color: rgb(255,0,0);" %)confirm with //enter//, click "<Call>"
375 -)))
376 -|(% colspan="1" %)(((
377 -Diode voltage
378 -)))|(% colspan="1" %)(((
379 -read out of diode
380 -)))|(% colspan="1" %)(((
381 -**~~ 0 !!! higher values indicate missaligned beam stop!!! 
382 -Call local contact
383 -**
384 -)))
385 -{{/layout-cell}}
386 -{{/layout-section}}
387 -
388 -{{layout-section ac:type="single"}}
389 -{{layout-cell}}
390 -----
391 -
392 -=== {{id name="14.SerialCrystallography:CFELTapeDrive-14.9Troubleshooting"/}}14.9 Troubleshooting ===
393 -
394 -* (((
395 -==== {{id name="14.SerialCrystallography:CFELTapeDrive-Tapestopped"/}}Tape stopped ====
396 -
397 -* Tape roll empty?
398 -* motor connected (icon green)?
399 -)))
400 -
401 -\\
402 -
403 -* (((
404 -==== {{id name="14.SerialCrystallography:CFELTapeDrive-Nozzlemotorsarenotresponding"/}}Nozzle motors are not responding ====
405 -
406 -* displayed in green=connected?
407 -* nozzle holder colliding somewhere?
408 -* motor limits reached? → position displayed in red
409 -)))
410 -
411 -\\
412 -
413 -* (((
414 -==== {{id name="14.SerialCrystallography:CFELTapeDrive-Sampleflowisnotrespondingtohigherpressure"/}}Sample flow is not responding to higher pressure ====
415 -
416 -* channel turned on?
417 -* Sample empty?
418 -* Nozzle or lines clogged? → flush with water, change nozzle
419 -)))
420 -{{/layout-cell}}
421 -{{/layout-section}}
422 -{{/layout}}
Confluence.Code.ConfluencePageClass[0]
Id
... ... @@ -1,1 +1,1 @@
1 -236824590
1 +235037709
URL
... ... @@ -1,1 +1,1 @@
1 -https://confluence.desy.de/spaces/11/pages/236824590/14. Serial Crystallography: CFEL TapeDrive
1 +https://confluence.desy.de/spaces/11/pages/235037709/14. Serial Crystallography: CFEL TapeDrive