... ... @@ -1,3 +1,5 @@
1 +[[image:attach:TapeDrive_beschriftet.PNG||height="400"]][[image:attach:TapeDrive scheme.png||height="400"]]
2 +
1 1  **14.1 Revert from Crystallography to TapeDrive (P11 staff/CFEL)**
2 2  
3 3  * remove vortex
... ... @@ -34,7 +34,8 @@
34 34  ** p11/motor/eh.09
35 35  * remote access to TapeDrive control laptop:
36 36  ** open terminal, execute command:          xfreerdp +clipboard /size:1920x1080 /u:Cfelusr /v:cfeld-pcx22611
37 - password: Sp!ceM3lange
39 + password: //written on control laptop
40 +//
38 38  * Cameras:
39 39  ** haspp11ethcam1 → detector tower camera to monitor nozzle
40 40  ** haspp11ethcam4 →  goniometer camera to monitor tape from above
... ... @@ -49,18 +49,64 @@
49 49  
50 50  **14.2 Sample characteristics and preparation**
51 51  
52 -suspension of microcrystals, **at least** 200 µL with 20% crystal density(more is always better), size range: 2-30 µm, **plus **few mL of mother liquor extra
55 +suspension of microcrystals, **at least** 200 µL with 20% crystal density(more is always better), size range: 2-30 µm, **tip:** bring some mother liquor for washing, filtering, diluting etc.
53 53  
54 54  Preparation:
55 55  
56 56  * filter through 30 µm cell filter (green) into Eppendorf or Falcon (original!) tube
57 57  
58 -[[image:url:https://www.sysmex.de/fileadmin/_processed_/a/d/csm_Sysmex_Partec_celltrics_filter_frei_20cacc97b9.png||width="500" height="334"]]
61 +[[image:attach:20211201_095534.jpg||height="250"]]
59 59  
60 60  \\
61 61  
65 +**14.3 Connecting sample to ElveFlow, sample exchange
66 +**
67 +
68 +[[image:attach:20211201_095033.jpg||thumbnail="true" height="250"]][[image:attach:20211201_095205.jpg||height="250"]][[image:attach:20211201_095232.jpg||height="250"]]
69 +
62 62  \\
63 63  
64 64  \\
65 65  
74 +**14.4 TapeDrive operation
75 +**
76 +
77 +The TapeDrive is controlled through kamzik. To open kamzik, start the server. A terminal will open, shortly after a small window.
78 +Now start the client. Same thing, a terminal will open, shortly after a small window.
79 +[[image:attach:screen.PNG||height="250"]][[image:attach:server client.PNG||height="250"]]
80 +
81 +* **Tape:**
82 +** Click on the 'main tape' icon
83 +** Change the velocity by typing anything from 0.3-3.0 mm/s
84 +** Common velocity is 1.0-1.5 mm/s
85 +** [[image:attach:kamzik_tape.PNG||height="400"]]
86 +
66 66  \\
88 +
89 +\\
90 +
91 +\\
92 +
93 +Once the hutch is interlocked, set remote control to the TapeDrive laptop by opening a terminal and executing the following command:
94 +
95 + xfreerdp +clipboard /size:1920x1080 /u:Cfelusr /v:cfeld-pcx22611
96 + password: //written on control laptop (changes every 6 months)
97 +//
98 +
99 +\\
100 +
101 +\\
102 +
103 +**14.5 Tape roll exchange**
104 +
105 +**[[image:attach:20211201_094111.jpg||height="250"]][[image:attach:20211201_094101.jpg||height="250"]]**
106 +
107 +\\
108 +
109 +\\
110 +
111 +\\
112 +
113 +\\
114 +
115 +\\
Confluence.Code.ConfluencePageClass[0]
Id
... ... @@ -1,1 +1,1 @@
1 -235034275
1 +235037525
URL
... ... @@ -1,1 +1,1 @@
1 -https://confluence.desy.de/spaces/11/pages/235034275/14. Serial Crystallography: CFEL TapeDrive
1 +https://confluence.desy.de/spaces/11/pages/235037525/14. Serial Crystallography: CFEL TapeDrive