214 214  * (% style="color: rgb(0,0,0);" %)set "Hole" to <Inner hole (largest)> (% style="color: rgb(255,0,0);" %)//Takes some minutes//
215 215  * (% style="color: rgb(0,0,0);" %)increase frequency by max. 10 Hz steps
216 216  ** (% style="color: rgb(0,0,0);" %)click arrow up 10x, "Set", arrow up 10x, "Set", .....
217 -* (% style="color: rgb(0,0,0);" %)//usually 133 Hz, results in exposure time of 3.7 ms//
217 +* (% style="color: rgb(0,0,0);" %)//usually 133 Hz, results in exposure time of 3.7 ms (technical exposure will always be 7.7 ms, since max. frame rate is 133 Hz
218 +//
218 218  {{/layout-cell}}
219 219  {{/layout-section}}
220 220  
... ... @@ -242,60 +242,35 @@
242 242  
243 243  (% style="color: rgb(255,0,0);" %)//always after entering the hutch!//
244 244  
245 -(% style="color: rgb(0,0,0);" %)Click "Safety" icon in kamzik.
246 +(% style="color: rgb(0,0,0);" %)Click "Run" icon in kamzik.
246 246  
247 247  * Put collimator in
248 -** Click "<Call>" next to //collimator_in//
249 +** Put detector distance 200 in //prepare_for_measurement (      )
250 +//
251 +** Click "<Call>" next to //prepare_for_measurement//
249 249  
250 250  \\
251 251  
252 252  * Open beam shutter BS1 in //CrystalControl //or// P11 XRayFeed //
253 -{{/layout-cell}}
254 254  
255 -{{layout-cell}}
256 -[[image:attach:kamzik_safety.PNG||height="400"]]
257 -{{/layout-cell}}
258 -{{/layout-section}}
257 +\\
259 259  
260 -{{layout-section ac:type="single"}}
261 -{{layout-cell}}
262 -----
259 +//**Set run parameters and go 🏃💥**//
263 263  
264 -Click "Start/Run" icon in kamzik.
261 +1. Set your desired No. of frames
262 +1. Match exposure time with chopper setting or set desired exposure time
263 +1. Check trigger setting and match to chopper setting
264 +1. To start the run, click "<Call>" next to //start_run (exposure    )//
265 +{{/layout-cell}}
265 265  
266 -[[image:attach:kamzik_run-tile.jpg||height="400"]]
267 -
268 -----
267 +{{layout-cell}}
268 +[[image:attach:Kamzik_run(2).PNG||height="400"]]
269 269  {{/layout-cell}}
270 270  {{/layout-section}}
271 271  
272 272  {{layout-section ac:type="two_equal"}}
273 273  {{layout-cell}}
274 -**Set run parameters and go 🏃💥**
275 -
276 -1. Set your desired No. of frames
277 -1. Match exposure time with chopper setting
278 -1. Check trigger setting and match to chopper setting
279 -1. Set detector distance to 200.0 mm, confirm with //enter// and click "<Call>" button next to it.
280 -1. To start the run, click "<Call>" next to //exposure time//
281 -
282 -\\
283 -
284 -\\
285 -
286 -**After data collection/Before entering the exp. hutch:**
287 -
288 -1. Go to "Safety" in kamzik.
289 -1. Click "<Call>" next to //prepare_to_open_hutch. //This will automatically:
290 -11. Drive the collimator out
291 -11. Drive the detector back to 1500 mm
292 -1. Close beam shutter BS1 and break door interlock in //P11 XRayFeed //(monitor top left).
293 -
294 -[[image:attach:Screenshot from 2022-04-04 19-36-29.png||height="400"]]
295 -{{/layout-cell}}
296 -
297 -{{layout-cell}}
298 -(% class="wrapped relative-table" style="width: 100.0%;" %)
274 +(% class="relative-table" %)
299 299  |=(((
300 300  Parameter
301 301  )))|=(((
... ... @@ -319,7 +319,7 @@
319 319  
320 320  \\
321 321  )))|(% colspan="1" %)(((
322 -chopper at 133 Hz → 3.7 ms
298 +chopper at 133 Hz → 7.7 ms (opening time will be 3.7)
323 323  
324 324  chopper at 66 Hz   → 7.7 ms
325 325  
... ... @@ -349,6 +349,18 @@
349 349  200.0 mm   (% style="color: rgb(255,0,0);" %)confirm with //enter//, click "<Call>"
350 350  )))
351 351  {{/layout-cell}}
328 +
329 +{{layout-cell}}
330 +**After data collection/Before entering the exp. hutch:**
331 +
332 +1. Go to "Safety" in kamzik.
333 +1. Click "<Call>" next to //prepare_to_open_hutch. //This will automatically:
334 +11. Drive the collimator out
335 +11. Drive the detector back to 1500 mm
336 +1. Close beam shutter BS1 and break door interlock in //P11 XRayFeed //(monitor top left).
337 +
338 +[[image:attach:Screenshot from 2022-04-04 19-36-29.png||height="400"]]
339 +{{/layout-cell}}
352 352  {{/layout-section}}
353 353  
354 354  {{layout-section ac:type="single"}}