... ... @@ -1,7 +1,7 @@
1 1  {{layout}}
2 2  {{layout-section ac:type="single"}}
3 3  {{layout-cell}}
4 -[[image:attach:TapeDrive_beschriftet-tile.jpg||height="400"]]
4 +[[image:attach:TapeDrive_beschriftet-tile.jpg||height="250"]]
5 5  
6 6  \\
7 7  
... ... @@ -13,11 +13,11 @@
13 13  
14 14  == {{id name="14.SerialCrystallography:CFELTapeDrive-14.1Samplecharacteristicsandpreparation"/}}**14.1 Sample characteristics and preparation** ==
15 15  
16 -suspension of microcrystals, **at least** 200 µL with 20% crystal density (more is always better), size range: 2-30 µm, **tip:** bring some mother liquor for washing, filtering, diluting etc.
16 +suspension of microcrystals, **at least** 200 µL with 20% crystal density(more is always better), size range: 2-30 µm, **tip:** bring some mother liquor for washing, filtering, diluting etc.
17 17  
18 18  Preparation:
19 19  
20 -* filter through 30 µm cell filter (green) or 50 µm (yellow) into Eppendorf or Falcon (original!) tube
20 +* filter through 30 µm cell filter (green) into Eppendorf or Falcon (original!) tube
21 21  
22 22  [[image:attach:20211201_095534.jpg||height="250"]]
23 23  
... ... @@ -57,7 +57,7 @@
57 57  
58 58  ==== {{id name="14.SerialCrystallography:CFELTapeDrive-Tape"/}}**Tape** ====
59 59  
60 -* Click the 'main tape' icon
60 +* Click on the 'main tape' icon
61 61  * Change the velocity by typing anything from **0.3-3.0 mm/s**
62 62  * Common velocity is 1.0-1.5 mm/s
63 63  * kamzik displays the used tape in m and %. At 98% the tape will automatically stop since the rolls need to be exchanged at this point.
... ... @@ -66,7 +66,7 @@
66 66  
67 67  ==== {{id name="14.SerialCrystallography:CFELTapeDrive-Samplenozzle"/}}**Sample nozzle** ====
68 68  
69 -* Click on the 'nozzle' icon.
69 +* Click an the 'nozzle' icon.
70 70  * (((
71 71  The nozzle position can be adjusted in xyz by clicking on the arrows up and down. (Double arrow = 10 steps). Step size can be varied, 1mm steps are rather big and should be used carefully.
72 72  The absolute position is **not calibrated** and therefore changes after each restart of kamzik.
... ... @@ -156,7 +156,7 @@
156 156  * Get a new tape roll (bottom left on the cart). **Polypropylen 15 µm**
157 157  * Cut the first cm of the new tape roll, directly behind the black sticky part.
158 158  * Place new roll on roll holder and use sticky tape to attach the end of the old tape to the beginning of the new roll.
159 -* Open magnetic tape stabilizer on the **left side**. (% style="color: rgb(255,0,0);" %)//!!!BEAMSTOP!!!//
159 +* Open magnetic tape stabilizer on the **left side**. //!!!BEAMSTOP!!!//
160 160  * Start the tape (3mm/s). Once the connetion side has arrived at the collector roll, stop the tape (0mm/s).
161 161  * Again, cut the tape directly behind the connection side. Replace the the old roll with the **empty **one. Old roll can be discarded.
162 162  * Attach new tape to the empty roll by using some sticky tape.
... ... @@ -214,13 +214,14 @@
214 214  * (% style="color: rgb(0,0,0);" %)set "Hole" to <Inner hole (largest)> (% style="color: rgb(255,0,0);" %)//Takes some minutes//
215 215  * (% style="color: rgb(0,0,0);" %)increase frequency by max. 10 Hz steps
216 216  ** (% style="color: rgb(0,0,0);" %)click arrow up 10x, "Set", arrow up 10x, "Set", .....
217 -* (% style="color: rgb(0,0,0);" %)//usually 133 Hz, results in exposure time of 3.7 ms (technical exposure will always be 7.7 ms, since max. frame rate is 133 Hz
218 -//
217 +* (% style="color: rgb(0,0,0);" %)//usually 133 Hz, results in exposure time of 3.7 ms//
219 219  {{/layout-cell}}
220 220  {{/layout-section}}
221 221  
222 222  {{layout-section ac:type="single"}}
223 223  {{layout-cell}}
223 +----
224 +
224 224  == {{id name="14.SerialCrystallography:CFELTapeDrive-14.6Datacollection"/}}**14.6 Data collection** ==
225 225  
226 226  Once the hutch is interlocked, set remote control to the TapeDrive laptop by opening a terminal and executing the following command:
... ... @@ -231,68 +231,56 @@
231 231   password: written on control laptop (changes every 6 months)
232 232  )))
233 233  
234 - //>>usually the terminal is open, just repeat previous command with arrow up, enter and type in the password//
235 + //>>usually the terminal is open, just repeat previous command with arrow up, enter and type in the password//
235 235  
236 236  \\
238 +
239 +Click "Start" icon in kamzik.
240 +
241 +[[image:attach:kamzik_run-tile.jpg||height="400"]]
242 +
243 +----
237 237  {{/layout-cell}}
238 238  {{/layout-section}}
239 239  
240 -{{layout-section ac:type="two_left_sidebar"}}
247 +{{layout-section ac:type="two_equal"}}
241 241  {{layout-cell}}
242 242  **Check list:**
243 243  
244 244  (% style="color: rgb(255,0,0);" %)//always after entering the hutch!//
245 245  
246 -(% style="color: rgb(0,0,0);" %)Click "Run" icon in kamzik.
253 +* Open BS1  (//CrystalControl)//
254 +* Put collimator in (//CrystalControl)//
255 +* Check beamstop position = 27 000 (//Atk panel p11/motor/eh.3.09)//
256 +* Open fast shutter (//CrystalControl)//
257 +* Check diode voltage ~~ 0.0 (//kamzik Run/Diode voltage)//
258 +* Close fast shutter (//CrystalControl)//
247 247  
248 -* Put collimator in
249 -** Put detector distance 200 in //prepare_for_measurement (      )
250 -//
251 -** Click "<Call>" next to //prepare_for_measurement//
252 -
253 253  \\
254 254  
255 -* Open beam shutter BS1 in //CrystalControl //or// P11 XRayFeed //
256 -{{/layout-cell}}
257 -
258 -{{layout-cell}}
259 -[[image:attach:kamzik_safety.PNG||height="400"]]
260 -{{/layout-cell}}
261 -{{/layout-section}}
262 -
263 -{{layout-section ac:type="single"}}
264 -{{layout-cell}}
265 -----
266 -
267 -[[image:attach:kamzik_run-tile.jpg||height="400"]]
268 -
269 -----
270 -{{/layout-cell}}
271 -{{/layout-section}}
272 -
273 -{{layout-section ac:type="two_equal"}}
274 -{{layout-cell}}
275 275  **Set run parameters and go 🏃💥**
276 276  
277 277  1. Set your desired No. of frames
278 -1. Match exposure time with chopper setting or set desired exposure time
265 +1. Match exposure time with chopper setting
279 279  1. Check trigger setting and match to chopper setting
280 -1. To start the run, click "<Call>" next to //start_run (exposure    )
281 -//
267 +1. Set detector distance to 200.0 mm, confirm with //enter// and click "<Call>" button next to it.
268 +1. To start the run, click "<Call>" next to //exposure time//
282 282  
283 283  \\
284 284  
285 -\\
272 +(% style="color: rgb(153,51,102);" %)**//Current bug~://**
286 286  
287 -**After data collection/Before entering the exp. hutch:**
274 +(% style="color: rgb(0,0,0);" %)You need to "Stop" run in kamzik manually after all images have been acquired. This moves the shield up and closes the fast shutter. Afterwards you can start a new run by clicking "<Call>".
288 288  
289 -1. Go to "Safety" in kamzik.
290 -1. Click "<Call>" next to //prepare_to_open_hutch. //This will automatically:
291 -11. Drive the collimator out
292 -11. Drive the detector back to 1500 mm
293 -1. Close beam shutter BS1 and break door interlock in //P11 XRayFeed //(monitor top left).
294 294  
295 -[[image:attach:Screenshot from 2022-04-04 19-36-29.png||height="400"]]
277 +(% style="color: rgb(0,0,0);" %)Check progress in terminal with:
278 +
279 +(% class="code" %)
280 +(((
281 + cd /gpfs/current/raw/scan_frames
282 + watch "ls -lrt <run number>/ | tail"
283 +\\\\\\\\
284 +)))
296 296  {{/layout-cell}}
297 297  
298 298  {{layout-cell}}
... ... @@ -320,7 +320,7 @@
320 320  
321 321  \\
322 322  )))|(% colspan="1" %)(((
323 -chopper at 133 Hz → 7.7 ms (opening time will be 3.7)
312 +chopper at 133 Hz → 3.7 ms
324 324  
325 325  chopper at 66 Hz   → 7.7 ms
326 326  
... ... @@ -349,6 +349,15 @@
349 349  )))|(% colspan="1" %)(((
350 350  200.0 mm   (% style="color: rgb(255,0,0);" %)confirm with //enter//, click "<Call>"
351 351  )))
341 +|(% colspan="1" %)(((
342 +Diode voltage
343 +)))|(% colspan="1" %)(((
344 +read out of diode
345 +)))|(% colspan="1" %)(((
346 +**~~ 0 !!! higher values indicate missaligned beam stop!!! 
347 +Call local contact
348 +**
349 +)))
352 352  {{/layout-cell}}
353 353  {{/layout-section}}
354 354  
... ... @@ -384,41 +384,16 @@
384 384  * Sample empty?
385 385  * Nozzle or lines clogged? → flush with water, change nozzle
386 386  )))
387 -
388 -\\
389 -
390 390  * (((
391 391  ==== {{id name="14.SerialCrystallography:CFELTapeDrive-Simplonfilewriter/eigerstreamissues"/}}Simplon filewriter/eiger stream issues ====
392 392  
393 393  * First, restart kamzik (server+client) and see if this solves the issue
394 394  * If not, restart the Simplon server (in Astor → haspp11exp03, right click on Simplon/EH.1 and restart)
395 -* (((
396 -If nothing works, kill the Simplon process and then start the server again:* (((
397 -(% class="code" %)
398 -(((
399 -ssh haspp11exp03
390 +* If nothing works, kill the Simplon process and then start the server again:
391 +** ssh haspp11exp03
392 +** ps -ef | grep Simplon
393 +** kill -9 simplon_process_id
400 400  )))
401 -)))
402 -* (((
403 -(% class="code" %)
404 -(((
405 -ps -ef | grep Simplon
406 -)))
407 -)))
408 -* (((
409 -(% class="code" %)
410 -(((
411 -kill -9 simplon_process_id
412 -)))
413 -)))
414 -)))
415 -)))
416 416  {{/layout-cell}}
417 417  {{/layout-section}}
418 -
419 -{{layout-section ac:type="single"}}
420 -{{layout-cell}}
421 -\\
422 -{{/layout-cell}}
423 -{{/layout-section}}
424 424  {{/layout}}