Last modified by chatzies on 2025-11-27 11:35

From version 67.1
edited by ahenkel
on 2022-05-07 18:33
Change comment: There is no comment for this version
To version 68.1
edited by ahenkel
on 2022-05-07 18:53
Change comment: There is no comment for this version

Summary

Details

Page properties
Content
... ... @@ -253,50 +253,25 @@
253 253  \\
254 254  
255 255  * Open beam shutter BS1 in //CrystalControl //or// P11 XRayFeed //
256 -{{/layout-cell}}
257 257  
258 -{{layout-cell}}
259 -[[image:attach:kamzik_safety.PNG||height="400"]]
260 -{{/layout-cell}}
261 -{{/layout-section}}
257 +\\
262 262  
263 -{{layout-section ac:type="single"}}
264 -{{layout-cell}}
265 -----
259 +//**Set run parameters and go 🏃💥**//
266 266  
267 -[[image:attach:kamzik_run-tile.jpg||height="400"]]
268 -
269 -----
270 -{{/layout-cell}}
271 -{{/layout-section}}
272 -
273 -{{layout-section ac:type="two_equal"}}
274 -{{layout-cell}}
275 -**Set run parameters and go 🏃💥**
276 -
277 277  1. Set your desired No. of frames
278 278  1. Match exposure time with chopper setting or set desired exposure time
279 279  1. Check trigger setting and match to chopper setting
280 -1. To start the run, click "<Call>" next to //start_run (exposure    )
281 -//
264 +1. To start the run, click "<Call>" next to //start_run (exposure    )//
265 +{{/layout-cell}}
282 282  
283 -\\
284 -
285 -\\
286 -
287 -**After data collection/Before entering the exp. hutch:**
288 -
289 -1. Go to "Safety" in kamzik.
290 -1. Click "<Call>" next to //prepare_to_open_hutch. //This will automatically:
291 -11. Drive the collimator out
292 -11. Drive the detector back to 1500 mm
293 -1. Close beam shutter BS1 and break door interlock in //P11 XRayFeed //(monitor top left).
294 -
295 -[[image:attach:Screenshot from 2022-04-04 19-36-29.png||height="400"]]
267 +{{layout-cell}}
268 +[[image:attach:Kamzik_run(2).PNG||height="400"]]
296 296  {{/layout-cell}}
270 +{{/layout-section}}
297 297  
272 +{{layout-section ac:type="two_equal"}}
298 298  {{layout-cell}}
299 -(% class="wrapped relative-table" style="width: 100.0%;" %)
274 +(% class="relative-table" %)
300 300  |=(((
301 301  Parameter
302 302  )))|=(((
... ... @@ -350,6 +350,18 @@
350 350  200.0 mm   (% style="color: rgb(255,0,0);" %)confirm with //enter//, click "<Call>"
351 351  )))
352 352  {{/layout-cell}}
328 +
329 +{{layout-cell}}
330 +**After data collection/Before entering the exp. hutch:**
331 +
332 +1. Go to "Safety" in kamzik.
333 +1. Click "<Call>" next to //prepare_to_open_hutch. //This will automatically:
334 +11. Drive the collimator out
335 +11. Drive the detector back to 1500 mm
336 +1. Close beam shutter BS1 and break door interlock in //P11 XRayFeed //(monitor top left).
337 +
338 +[[image:attach:Screenshot from 2022-04-04 19-36-29.png||height="400"]]
339 +{{/layout-cell}}
353 353  {{/layout-section}}
354 354  
355 355  {{layout-section ac:type="single"}}
Confluence.Code.ConfluencePageClass[0]
Id
... ... @@ -1,1 +1,1 @@
1 -254729368
1 +254729370
URL
... ... @@ -1,1 +1,1 @@
1 -https://confluence.desy.de/spaces/11/pages/254729368/14. Serial Crystallography: CFEL TapeDrive
1 +https://confluence.desy.de/spaces/11/pages/254729370/14. Serial Crystallography: CFEL TapeDrive