... ... @@ -214,8 +214,7 @@
214 214  * (% style="color: rgb(0,0,0);" %)set "Hole" to <Inner hole (largest)> (% style="color: rgb(255,0,0);" %)//Takes some minutes//
215 215  * (% style="color: rgb(0,0,0);" %)increase frequency by max. 10 Hz steps
216 216  ** (% style="color: rgb(0,0,0);" %)click arrow up 10x, "Set", arrow up 10x, "Set", .....
217 -* (% style="color: rgb(0,0,0);" %)//usually 133 Hz, results in exposure time of 3.7 ms (technical exposure will always be 7.7 ms, since max. frame rate is 133 Hz
218 -//
217 +* (% style="color: rgb(0,0,0);" %)//usually 133 Hz, results in exposure time of 3.7 ms//
219 219  {{/layout-cell}}
220 220  {{/layout-section}}
221 221  
... ... @@ -243,35 +243,60 @@
243 243  
244 244  (% style="color: rgb(255,0,0);" %)//always after entering the hutch!//
245 245  
246 -(% style="color: rgb(0,0,0);" %)Click "Run" icon in kamzik.
245 +(% style="color: rgb(0,0,0);" %)Click "Safety" icon in kamzik.
247 247  
248 248  * Put collimator in
249 -** Put detector distance 200 in //prepare_for_measurement (      )
250 -//
251 -** Click "<Call>" next to //prepare_for_measurement//
248 +** Click "<Call>" next to //collimator_in//
252 252  
253 253  \\
254 254  
255 255  * Open beam shutter BS1 in //CrystalControl //or// P11 XRayFeed //
253 +{{/layout-cell}}
256 256  
257 -\\
258 -
259 -//**Set run parameters and go 🏃💥**//
260 -
261 -1. Set your desired No. of frames
262 -1. Match exposure time with chopper setting or set desired exposure time
263 -1. Check trigger setting and match to chopper setting
264 -1. To start the run, click "<Call>" next to //start_run (exposure    )//
255 +{{layout-cell}}
256 +[[image:attach:kamzik_safety.PNG||height="400"]]
265 265  {{/layout-cell}}
258 +{{/layout-section}}
266 266  
260 +{{layout-section ac:type="single"}}
267 267  {{layout-cell}}
268 -[[image:attach:Kamzik_run(2).PNG||height="400"]]
262 +----
263 +
264 +Click "Start/Run" icon in kamzik.
265 +
266 +[[image:attach:kamzik_run-tile.jpg||height="400"]]
267 +
268 +----
269 269  {{/layout-cell}}
270 270  {{/layout-section}}
271 271  
272 272  {{layout-section ac:type="two_equal"}}
273 273  {{layout-cell}}
274 -(% class="relative-table" %)
274 +**Set run parameters and go 🏃💥**
275 +
276 +1. Set your desired No. of frames
277 +1. Match exposure time with chopper setting
278 +1. Check trigger setting and match to chopper setting
279 +1. Set detector distance to 200.0 mm, confirm with //enter// and click "<Call>" button next to it.
280 +1. To start the run, click "<Call>" next to //exposure time//
281 +
282 +\\
283 +
284 +\\
285 +
286 +**After data collection/Before entering the exp. hutch:**
287 +
288 +1. Go to "Safety" in kamzik.
289 +1. Click "<Call>" next to //prepare_to_open_hutch. //This will automatically:
290 +11. Drive the collimator out
291 +11. Drive the detector back to 1500 mm
292 +1. Close beam shutter BS1 and break door interlock in //P11 XRayFeed //(monitor top left).
293 +
294 +[[image:attach:Screenshot from 2022-04-04 19-36-29.png||height="400"]]
295 +{{/layout-cell}}
296 +
297 +{{layout-cell}}
298 +(% class="wrapped relative-table" style="width: 100.0%;" %)
275 275  |=(((
276 276  Parameter
277 277  )))|=(((
... ... @@ -295,7 +295,7 @@
295 295  
296 296  \\
297 297  )))|(% colspan="1" %)(((
298 -chopper at 133 Hz → 7.7 ms (opening time will be 3.7)
322 +chopper at 133 Hz → 3.7 ms
299 299  
300 300  chopper at 66 Hz   → 7.7 ms
301 301  
... ... @@ -325,18 +325,6 @@
325 325  200.0 mm   (% style="color: rgb(255,0,0);" %)confirm with //enter//, click "<Call>"
326 326  )))
327 327  {{/layout-cell}}
328 -
329 -{{layout-cell}}
330 -**After data collection/Before entering the exp. hutch:**
331 -
332 -1. Go to "Safety" in kamzik.
333 -1. Click "<Call>" next to //prepare_to_open_hutch. //This will automatically:
334 -11. Drive the collimator out
335 -11. Drive the detector back to 1500 mm
336 -1. Close beam shutter BS1 and break door interlock in //P11 XRayFeed //(monitor top left).
337 -
338 -[[image:attach:Screenshot from 2022-04-04 19-36-29.png||height="400"]]
339 -{{/layout-cell}}
340 340  {{/layout-section}}
341 341  
342 342  {{layout-section ac:type="single"}}