... ... @@ -214,15 +214,12 @@
214 214  * (% style="color: rgb(0,0,0);" %)set "Hole" to <Inner hole (largest)> (% style="color: rgb(255,0,0);" %)//Takes some minutes//
215 215  * (% style="color: rgb(0,0,0);" %)increase frequency by max. 10 Hz steps
216 216  ** (% style="color: rgb(0,0,0);" %)click arrow up 10x, "Set", arrow up 10x, "Set", .....
217 -* (% style="color: rgb(0,0,0);" %)//usually 133 Hz, results in exposure time of 3.7 ms (technical exposure will always be 7.7 ms, since max. frame rate is 133 Hz
218 -//
217 +* (% style="color: rgb(0,0,0);" %)//usually 133 Hz, results in exposure time of 3.7 ms//
219 219  {{/layout-cell}}
220 220  {{/layout-section}}
221 221  
222 222  {{layout-section ac:type="single"}}
223 223  {{layout-cell}}
224 -----
225 -
226 226  == {{id name="14.SerialCrystallography:CFELTapeDrive-14.6Datacollection"/}}**14.6 Data collection** ==
227 227  
228 228  Once the hutch is interlocked, set remote control to the TapeDrive laptop by opening a terminal and executing the following command:
... ... @@ -245,35 +245,60 @@
245 245  
246 246  (% style="color: rgb(255,0,0);" %)//always after entering the hutch!//
247 247  
248 -(% style="color: rgb(0,0,0);" %)Click "Run" icon in kamzik.
245 +(% style="color: rgb(0,0,0);" %)Click "Safety" icon in kamzik.
249 249  
250 250  * Put collimator in
251 -** Put detector distance 200 in //prepare_for_measurement (      )
252 -//
253 -** Click "<Call>" next to //prepare_for_measurement//
248 +** Click "<Call>" next to //collimator_in//
254 254  
255 255  \\
256 256  
257 257  * Open beam shutter BS1 in //CrystalControl //or// P11 XRayFeed //
253 +{{/layout-cell}}
258 258  
259 -\\
260 -
261 -//**Set run parameters and go 🏃💥**//
262 -
263 -1. Set your desired No. of frames
264 -1. Match exposure time with chopper setting or set desired exposure time
265 -1. Check trigger setting and match to chopper setting
266 -1. To start the run, click "<Call>" next to //start_run (exposure    )//
255 +{{layout-cell}}
256 +[[image:attach:kamzik_safety.PNG||height="400"]]
267 267  {{/layout-cell}}
258 +{{/layout-section}}
268 268  
260 +{{layout-section ac:type="single"}}
269 269  {{layout-cell}}
270 -[[image:attach:Kamzik_run(2).PNG]]
262 +----
263 +
264 +Click "Start/Run" icon in kamzik.
265 +
266 +[[image:attach:kamzik_run-tile.jpg||height="400"]]
267 +
268 +----
271 271  {{/layout-cell}}
272 272  {{/layout-section}}
273 273  
274 274  {{layout-section ac:type="two_equal"}}
275 275  {{layout-cell}}
276 -(% class="relative-table" %)
274 +**Set run parameters and go 🏃💥**
275 +
276 +1. Set your desired No. of frames
277 +1. Match exposure time with chopper setting
278 +1. Check trigger setting and match to chopper setting
279 +1. Set detector distance to 200.0 mm, confirm with //enter// and click "<Call>" button next to it.
280 +1. To start the run, click "<Call>" next to //exposure time//
281 +
282 +\\
283 +
284 +\\
285 +
286 +**After data collection/Before entering the exp. hutch:**
287 +
288 +1. Go to "Safety" in kamzik.
289 +1. Click "<Call>" next to //prepare_to_open_hutch. //This will automatically:
290 +11. Drive the collimator out
291 +11. Drive the detector back to 1500 mm
292 +1. Close beam shutter BS1 and break door interlock in //P11 XRayFeed //(monitor top left).
293 +
294 +[[image:attach:Screenshot from 2022-04-04 19-36-29.png||height="400"]]
295 +{{/layout-cell}}
296 +
297 +{{layout-cell}}
298 +(% class="wrapped relative-table" style="width: 100.0%;" %)
277 277  |=(((
278 278  Parameter
279 279  )))|=(((
... ... @@ -297,7 +297,7 @@
297 297  
298 298  \\
299 299  )))|(% colspan="1" %)(((
300 -chopper at 133 Hz → 7.7 ms (opening time will be 3.7)
322 +chopper at 133 Hz → 3.7 ms
301 301  
302 302  chopper at 66 Hz   → 7.7 ms
303 303  
... ... @@ -326,33 +326,18 @@
326 326  )))|(% colspan="1" %)(((
327 327  200.0 mm   (% style="color: rgb(255,0,0);" %)confirm with //enter//, click "<Call>"
328 328  )))
351 +|(% colspan="1" %)(((
352 +Diode voltage
353 +)))|(% colspan="1" %)(((
354 +read out of diode
355 +)))|(% colspan="1" %)(((
356 +**~~ 0 !!! higher values indicate missaligned beam stop!!! 
357 +Call local contact
358 +**
359 +)))
329 329  {{/layout-cell}}
330 -
331 -{{layout-cell}}
332 -\\
333 -{{/layout-cell}}
334 334  {{/layout-section}}
335 335  
336 -{{layout-section ac:type="two_equal"}}
337 -{{layout-cell}}
338 -**After data collection/Before entering the exp. hutch:**
339 -
340 -1. Go to "Safety" in kamzik.
341 -1. Click "<Call>" next to //prepare_to_open_hutch. //This will automatically:
342 -11. Drive the collimator out
343 -11. Drive the detector back to 1500 mm
344 -1. Close beam shutter BS1 and break door interlock in //P11 XRayFeed //(monitor top left).
345 -
346 -[[image:attach:Screenshot from 2022-04-04 19-36-29.png||height="400"]]
347 -{{/layout-cell}}
348 -
349 -{{layout-cell}}
350 -[[image:attach:safety_kamzik.PNG]]
351 -
352 -\\
353 -{{/layout-cell}}
354 -{{/layout-section}}
355 -
356 356  {{layout-section ac:type="single"}}
357 357  {{layout-cell}}
358 358  ----
Confluence.Code.ConfluencePageClass[0]
Id
... ... @@ -1,1 +1,1 @@
1 -254729371
1 +252826012
URL
... ... @@ -1,1 +1,1 @@
1 -https://confluence.desy.de/spaces/11/pages/254729371/14. Serial Crystallography: CFEL TapeDrive
1 +https://confluence.desy.de/spaces/11/pages/252826012/14. Serial Crystallography: CFEL TapeDrive