... ... @@ -1,7 +1,7 @@
1 1  {{layout}}
2 2  {{layout-section ac:type="single"}}
3 3  {{layout-cell}}
4 -[[image:attach:TapeDrive_beschriftet-tile.jpg||height="400"]]
4 +[[image:attach:TapeDrive_beschriftet-tile.jpg||height="250"]]
5 5  
6 6  \\
7 7  
... ... @@ -9,22 +9,18 @@
9 9  
10 10  \\
11 11  
12 -----
13 -
14 14  == {{id name="14.SerialCrystallography:CFELTapeDrive-14.1Samplecharacteristicsandpreparation"/}}**14.1 Sample characteristics and preparation** ==
15 15  
16 -suspension of microcrystals, **at least** 200 µL with 20% crystal density (more is always better), size range: 2-30 µm, **tip:** bring some mother liquor for washing, filtering, diluting etc.
14 +suspension of microcrystals, **at least** 200 µL with 20% crystal density(more is always better), size range: 2-30 µm, **tip:** bring some mother liquor for washing, filtering, diluting etc.
17 17  
18 18  Preparation:
19 19  
20 -* filter through 30 µm cell filter (green) or 50 µm (yellow) into Eppendorf or Falcon (original!) tube
18 +* filter through 30 µm cell filter (green) into Eppendorf or Falcon (original!) tube
21 21  
22 22  [[image:attach:20211201_095534.jpg||height="250"]]
23 23  
24 24  \\
25 25  
26 -----
27 -
28 28  == {{id name="14.SerialCrystallography:CFELTapeDrive-14.2ConnectingsampletoElveFlow,sampleexchange"/}}**14.2 Connecting sample to ElveFlow, sample exchange
29 29  ** ==
30 30  
... ... @@ -44,8 +44,6 @@
44 44  
45 45  \\
46 46  
47 -----
48 -
49 49  == {{id name="14.SerialCrystallography:CFELTapeDrive-14.3TapeDriveoperation"/}}**14.3 TapeDrive operation
50 50  ** ==
51 51  
... ... @@ -57,7 +57,7 @@
57 57  
58 58  ==== {{id name="14.SerialCrystallography:CFELTapeDrive-Tape"/}}**Tape** ====
59 59  
60 -* Click the 'main tape' icon
54 +* Click on the 'main tape' icon
61 61  * Change the velocity by typing anything from **0.3-3.0 mm/s**
62 62  * Common velocity is 1.0-1.5 mm/s
63 63  * kamzik displays the used tape in m and %. At 98% the tape will automatically stop since the rolls need to be exchanged at this point.
... ... @@ -66,7 +66,7 @@
66 66  
67 67  ==== {{id name="14.SerialCrystallography:CFELTapeDrive-Samplenozzle"/}}**Sample nozzle** ====
68 68  
69 -* Click on the 'nozzle' icon.
63 +* Click an the 'nozzle' icon.
70 70  * (((
71 71  The nozzle position can be adjusted in xyz by clicking on the arrows up and down. (Double arrow = 10 steps). Step size can be varied, 1mm steps are rather big and should be used carefully.
72 72  The absolute position is **not calibrated** and therefore changes after each restart of kamzik.
... ... @@ -145,10 +145,8 @@
145 145  
146 146  \\
147 147  
148 -----
142 +== {{id name="14.SerialCrystallography:CFELTapeDrive-14.5Taperollexchange"/}}**14.5 Tape roll exchange** ==
149 149  
150 -== {{id name="14.SerialCrystallography:CFELTapeDrive-14.4Taperollexchange"/}}**14.4 Tape roll exchange** ==
151 -
152 152  The tape stops automatically at 98% used length, if you need to change the rolls for orther reasons, stop the tape (velocity = 0.0 mm/s).
153 153  
154 154  * Remove the (almost) empty feeder roll by pulling it upwards.
... ... @@ -156,7 +156,7 @@
156 156  * Get a new tape roll (bottom left on the cart). **Polypropylen 15 µm**
157 157  * Cut the first cm of the new tape roll, directly behind the black sticky part.
158 158  * Place new roll on roll holder and use sticky tape to attach the end of the old tape to the beginning of the new roll.
159 -* Open magnetic tape stabilizer on the **left side**. (% style="color: rgb(255,0,0);" %)//!!!BEAMSTOP!!!//
151 +* Open magnetic tape stabilizer on the **left side**. //!!!BEAMSTOP!!!//
160 160  * Start the tape (3mm/s). Once the connetion side has arrived at the collector roll, stop the tape (0mm/s).
161 161  * Again, cut the tape directly behind the connection side. Replace the the old roll with the **empty **one. Old roll can be discarded.
162 162  * Attach new tape to the empty roll by using some sticky tape.
... ... @@ -170,17 +170,15 @@
170 170  
171 171  \\
172 172  
173 -**
174 -**
165 +=== {{id name="14.SerialCrystallography:CFELTapeDrive-"/}}**
166 +** ===
175 175  {{/layout-cell}}
176 176  {{/layout-section}}
177 177  
178 178  {{layout-section ac:type="single"}}
179 179  {{layout-cell}}
180 -----
172 +== {{id name="14.SerialCrystallography:CFELTapeDrive-14.7Chopper"/}}**14.7 Chopper** ==
181 181  
182 -== {{id name="14.SerialCrystallography:CFELTapeDrive-14.5Chopper"/}}**14.5 Chopper** ==
183 -
184 184  start chopper gui from terminal:
185 185  
186 186  (% class="code" %)
... ... @@ -191,7 +191,6 @@
191 191  (% class="code" %)
192 192  (((
193 193   python chopperControl.py
194 -\\\\
195 195  )))
196 196  {{/layout-cell}}
197 197  {{/layout-section}}
... ... @@ -214,17 +214,14 @@
214 214  * (% style="color: rgb(0,0,0);" %)set "Hole" to <Inner hole (largest)> (% style="color: rgb(255,0,0);" %)//Takes some minutes//
215 215  * (% style="color: rgb(0,0,0);" %)increase frequency by max. 10 Hz steps
216 216  ** (% style="color: rgb(0,0,0);" %)click arrow up 10x, "Set", arrow up 10x, "Set", .....
217 -* (% style="color: rgb(0,0,0);" %)//usually 133 Hz, results in exposure time of 3.7 ms (technical exposure will always be 7.7 ms, since max. frame rate is 133 Hz
218 -//
206 +* (% style="color: rgb(0,0,0);" %)//usually 133 Hz, results in exposure time of 3.7 ms//
219 219  {{/layout-cell}}
220 220  {{/layout-section}}
221 221  
222 222  {{layout-section ac:type="single"}}
223 223  {{layout-cell}}
224 -----
212 +== {{id name="14.SerialCrystallography:CFELTapeDrive-14.8Datacollection"/}}**14.8 Data collection** ==
225 225  
226 -== {{id name="14.SerialCrystallography:CFELTapeDrive-14.6Datacollection"/}}**14.6 Data collection** ==
227 -
228 228  Once the hutch is interlocked, set remote control to the TapeDrive laptop by opening a terminal and executing the following command:
229 229  
230 230  (% class="code" %)
... ... @@ -233,47 +233,60 @@
233 233   password: written on control laptop (changes every 6 months)
234 234  )))
235 235  
236 - //>>usually the terminal is open, just repeat previous command with arrow up, enter and type in the password//
222 + //>>usually the terminal is open, just repeat previous command with arrow up, enter and type in the password//
237 237  
238 238  \\
225 +
226 +Click "Start" icon in kamzik.
227 +
228 +[[image:attach:kamzik_run-tile.jpg||height="400"]]
229 +
230 +----
239 239  {{/layout-cell}}
240 240  {{/layout-section}}
241 241  
242 -{{layout-section ac:type="two_left_sidebar"}}
234 +{{layout-section ac:type="two_equal"}}
243 243  {{layout-cell}}
244 244  **Check list:**
245 245  
246 246  (% style="color: rgb(255,0,0);" %)//always after entering the hutch!//
247 247  
248 -(% style="color: rgb(0,0,0);" %)Click "Run" icon in kamzik.
240 +* Open BS1  (//CrystalControl)//
241 +* Put collimator in (//CrystalControl)//
242 +* Check beamstop position = 27 000 (//Atk panel p11/motor/eh.3.09)//
243 +* Open fast shutter (//CrystalControl)//
244 +* Check diode voltage ~~ 0.0 (//kamzik Run/Diode voltage)//
245 +* Close fast shutter (//CrystalControl)//
249 249  
250 -* Put collimator in
251 -** Put detector distance 200 in //prepare_for_measurement (      )
252 -//
253 -** Click "<Call>" next to //prepare_for_measurement//
254 -
255 255  \\
256 256  
257 -* Open beam shutter BS1 in //CrystalControl //or// P11 XRayFeed //
249 +**Set run parameters and go 🏃💥**
258 258  
251 +1. Set your desired No. of frames
252 +1. Match exposure time with chopper setting
253 +1. Check trigger setting and match to chopper setting
254 +1. Set detector distance to 200.0 mm, confirm with //enter// and click "<Call>" button next to it.
255 +1. To start the run, click "<Call>" next to //exposure time//
256 +
259 259  \\
260 260  
261 -//**Set run parameters and go 🏃💥**//
259 +(% style="color: rgb(153,51,102);" %)**//Current bug~://**
262 262  
263 -1. Set your desired No. of frames
264 -1. Match exposure time with chopper setting or set desired exposure time
265 -1. Check trigger setting and match to chopper setting
266 -1. To start the run, click "<Call>" next to //start_run (exposure    )//
267 -{{/layout-cell}}
261 +(% style="color: rgb(0,0,0);" %)You need to "Stop" run in kamzik manually after all images have been acquired. This moves the shield up and closes the fast shutter. Afterwards you can start a new run by clicking "<Call>".
268 268  
269 -{{layout-cell}}
270 -[[image:attach:Kamzik_run(2).PNG]]
263 +
264 +(% style="color: rgb(0,0,0);" %)Check progress in terminal with:
265 +
266 +(% class="code" %)
267 +(((
268 + cd /gpfs/current/raw/scan_frames
269 + watch "ls -lrt <run number>/ | tail"
270 +\\\\\\\\
271 +)))
271 271  {{/layout-cell}}
272 -{{/layout-section}}
273 273  
274 -{{layout-section ac:type="two_equal"}}
275 275  {{layout-cell}}
276 -(% class="relative-table wrapped" %)
275 +(% class="wrapped relative-table" style="width: 100.0%;" %)
277 277  |=(((
278 278  Parameter
279 279  )))|=(((
... ... @@ -297,7 +297,7 @@
297 297  
298 298  \\
299 299  )))|(% colspan="1" %)(((
300 -chopper at 133 Hz → 7.7 ms (opening time will be 3.7)
299 +chopper at 133 Hz → 3.7 ms
301 301  
302 302  chopper at 66 Hz   → 7.7 ms
303 303  
... ... @@ -326,42 +326,23 @@
326 326  )))|(% colspan="1" %)(((
327 327  200.0 mm   (% style="color: rgb(255,0,0);" %)confirm with //enter//, click "<Call>"
328 328  )))
328 +|(% colspan="1" %)(((
329 +Diode voltage
330 +)))|(% colspan="1" %)(((
331 +read out of diode
332 +)))|(% colspan="1" %)(((
333 +**~~ 0 !!! higher values indicate missaligned beam stop!!! 
334 +Call local contact
335 +**
336 +)))
329 329  {{/layout-cell}}
330 -
331 -{{layout-cell}}
332 -\\
333 -{{/layout-cell}}
334 334  {{/layout-section}}
335 335  
336 -{{layout-section ac:type="two_equal"}}
337 -{{layout-cell}}
338 -----
339 -
340 -**After data collection/Before entering the exp. hutch:**
341 -
342 -1. Go to "Safety" in kamzik.
343 -1. Click "<Call>" next to //prepare_to_open_hutch. //This will automatically:
344 -11. Drive the collimator out
345 -11. Drive the detector back to 1500 mm
346 -1. Close beam shutter BS1 and break door interlock in //P11 XRayFeed //(monitor top left).
347 -
348 -[[image:attach:Screenshot from 2022-04-04 19-36-29.png]]
349 -{{/layout-cell}}
350 -
351 -{{layout-cell}}
352 -----
353 -
354 -[[image:attach:safety_kamzik.PNG]]
355 -
356 -\\
357 -{{/layout-cell}}
358 -{{/layout-section}}
359 -
360 360  {{layout-section ac:type="single"}}
361 361  {{layout-cell}}
362 362  ----
363 363  
364 -== {{id name="14.SerialCrystallography:CFELTapeDrive-14.7Troubleshooting"/}}**14.7 Troubleshooting** ==
344 +== {{id name="14.SerialCrystallography:CFELTapeDrive-14.9Troubleshooting"/}}**14.9 Troubleshooting** ==
365 365  
366 366  * (((
367 367  ==== {{id name="14.SerialCrystallography:CFELTapeDrive-Tapestopped"/}}Tape stopped ====
... ... @@ -389,41 +389,6 @@
389 389  * Sample empty?
390 390  * Nozzle or lines clogged? → flush with water, change nozzle
391 391  )))
392 -
393 -\\
394 -
395 -* (((
396 -==== {{id name="14.SerialCrystallography:CFELTapeDrive-Simplonfilewriter/eigerstreamissues"/}}Simplon filewriter/eiger stream issues ====
397 -
398 -* First, restart kamzik (server+client) and see if this solves the issue
399 -* If not, restart the Simplon server (in Astor → haspp11exp03, right click on Simplon/EH.1 and restart)
400 -* (((
401 -If nothing works, kill the Simplon process and then start the server again:* (((
402 -(% class="code" %)
403 -(((
404 -ssh haspp11exp03
405 -)))
406 -)))
407 -* (((
408 -(% class="code" %)
409 -(((
410 -ps -ef | grep Simplon
411 -)))
412 -)))
413 -* (((
414 -(% class="code" %)
415 -(((
416 -kill -9 simplon_process_id
417 -)))
418 -)))
419 -)))
420 -)))
421 421  {{/layout-cell}}
422 422  {{/layout-section}}
423 -
424 -{{layout-section ac:type="single"}}
425 -{{layout-cell}}
426 -\\
427 -{{/layout-cell}}
428 -{{/layout-section}}
429 429  {{/layout}}