... ... @@ -1,1 +1,1 @@
1 -14. Serial Crystallography: CFEL TapeDrive
1 +10. Serial Crystallography: CFEL TapeDrive
Author
... ... @@ -1,1 +1,1 @@
1 -XWiki.ahenkel
1 +XWiki.tabermah
Content
... ... @@ -3,11 +3,9 @@
3 3  {{layout-cell}}
4 4  [[image:attach:TapeDrive_beschriftet-tile.jpg||height="400"]]
5 5  
6 -\\
7 7  
8 8  {{toc minLevel="2"/}}
9 9  
10 -\\
11 11  
12 12  ----
13 13  
... ... @@ -21,11 +21,12 @@
21 21  
22 22  [[image:attach:20211201_095534.jpg||height="250"]]
23 23  
24 -\\
25 25  
26 26  ----
27 27  
28 -== {{id name="14.SerialCrystallography:CFELTapeDrive-14.2ConnectingsampletoElveFlow,sampleexchange"/}}**14.2 Connecting sample to ElveFlow, sample exchange
25 +== {{id name="14.SerialCrystallography:CFELTapeDrive-14.2ConnectingsampletoElveFlow,sampleexchange"/}}**14.2 Connecting sample to ElveFlow, sample exchange** ==
26 +
27 +*
29 29  ** ==
30 30  
31 31  After filtration, screw the ElveFlow connector to your Eppendorf or Falcon tube. Make sure that the yellow-ish line goes to the bottom of your tube. You can adjust the length by unscrewing the blue fitting on top and pushing/pulling the yellow-ish line. **Don't forget to tigthen the blue fitting afterwards!**
... ... @@ -42,14 +42,15 @@
42 42  
43 43  [[image:attach:20211201_095232.jpg||height="250"]][[image:attach:20211201_095205.jpg||height="250"]]
44 44  
45 -\\
46 46  
47 47  ----
48 48  
49 -== {{id name="14.SerialCrystallography:CFELTapeDrive-14.3TapeDriveoperation"/}}**14.3 TapeDrive operation
47 +== {{id name="14.SerialCrystallography:CFELTapeDrive-14.3TapeDriveoperation"/}}**14.3 TapeDrive operation** ==
48 +
49 +*
50 50  ** ==
51 51  
52 -The TapeDrive is controlled through kamzik. To open kamzik, start the server. A terminal will open and shortly after a small window.
52 +The TapeDrive is controlled through kamzik. To open kamzik, start the server. A terminal will open and shortly after a small window.
53 53  Now start the client. Same thing, a terminal will open, shortly after a small window.
54 54  
55 55  Click 'Devices' in the **client session** to open the control window.
... ... @@ -61,7 +61,7 @@
61 61  * Change the velocity by typing anything from **0.3-3.0 mm/s**
62 62  * Common velocity is 1.0-1.5 mm/s
63 63  * kamzik displays the used tape in m and %. At 98% the tape will automatically stop since the rolls need to be exchanged at this point.
64 -\\
64 +
65 65  * [[image:attach:kamzik_tape.PNG||height="400"]]
66 66  
67 67  ==== {{id name="14.SerialCrystallography:CFELTapeDrive-Samplenozzle"/}}**Sample nozzle** ====
... ... @@ -72,16 +72,16 @@
72 72  The absolute position is **not calibrated** and therefore changes after each restart of kamzik.
73 73  
74 74  
75 -(% style="color: rgb(255,0,0);" %)**!!! **(% style="color: rgb(255, 0, 0); color: rgb(0, 0, 0)" %)**During data collection make sure in which direction you are moving, do not move nozzle into beam**(% style="color: rgb(255,0,0);" %)** !!!**(%%)
75 +(% style="color:#ff0000" %)**!!! **(% style="color:#000000" %)**During data collection make sure in which direction you are moving, do not move nozzle into beam**(% style="color:#ff0000" %)** !!!**(%%)
76 76  \\Adjust the nozzle position by eye until it nearly touches the tape and roughly meets the X-ray interaction point. Then, move to the control hutch to use the on-axis microscope for fine adjustment. Move the nozzle tip to the beam position (red X) and write down the absolut nozzle position. From there move the nozzle in +x to your desired distance.Different distances combined with the tape speed result in several delay times for time-resolved experiments.
77 77  
78 78  (% class="wrapped" %)
79 79  |=(((
80 -\\
80 +
81 81  )))|=(((
82 82  direction
83 83  )))|=(((
84 -recommended step size
84 +recommended step size
85 85  for visual adjustment
86 86  )))|=(% colspan="1" %)(((
87 87  recommended step size
... ... @@ -116,8 +116,8 @@
116 116  be careful with -z, tape could be pushed out of focus
117 117  )))
118 118  
119 -**
120 -\\**
119 +
120 +
121 121  )))
122 122  * [[image:attach:kamzik_nozzle.PNG||height="400"]]
123 123  
... ... @@ -125,25 +125,19 @@
125 125  
126 126  * The chiller circulates fluid through the gas nozzle.
127 127  
128 -\\
129 129  
130 130  ==== {{id name="14.SerialCrystallography:CFELTapeDrive-Gasnozzle"/}}**Gas nozzle** ====
131 131  
132 -\\
133 133  
134 134  ==== {{id name="14.SerialCrystallography:CFELTapeDrive-IRcamera"/}}**IR camera** ====
135 135  
136 -\\
137 137  
138 138  Once the hutch is interlocked, set remote control to the TapeDrive laptop by opening a terminal and executing the following command:
139 139  
140 140   xfreerdp +clipboard /size:1920x1080 /u:Cfelusr /v:cfeld-pcx22611
141 - password: //written on control laptop (changes every 6 months)
142 -//
138 + password: //written on control laptop (changes every 6 months)//
143 143  
144 -\\
145 145  
146 -\\
147 147  
148 148  ----
149 149  
... ... @@ -156,7 +156,7 @@
156 156  * Get a new tape roll (bottom left on the cart). **Polypropylen 15 µm**
157 157  * Cut the first cm of the new tape roll, directly behind the black sticky part.
158 158  * Place new roll on roll holder and use sticky tape to attach the end of the old tape to the beginning of the new roll.
159 -* Open magnetic tape stabilizer on the **left side**. (% style="color: rgb(255,0,0);" %)//!!!BEAMSTOP!!!//
153 +* Open magnetic tape stabilizer on the **left side**. (% style="color:#ff0000" %)//!!!BEAMSTOP!!!//
160 160  * Start the tape (3mm/s). Once the connetion side has arrived at the collector roll, stop the tape (0mm/s).
161 161  * Again, cut the tape directly behind the connection side. Replace the the old roll with the **empty **one. Old roll can be discarded.
162 162  * Attach new tape to the empty roll by using some sticky tape.
... ... @@ -163,15 +163,11 @@
163 163  * Close tape stabilizer.
164 164  * Reset tape counter in kamzik by clicking "<Call>" in "Main-Tape".
165 165  
166 -**[[image:attach:20211201_094101.jpg||height="250"]][[image:attach:20211201_094111.jpg||height="250"]][[image:attach:tape reset.PNG||height="250"]]
167 -**
160 +**[[image:attach:20211201_094101.jpg||height="250"]][[image:attach:20211201_094111.jpg||height="250"]][[image:attach:tape reset.PNG||height="250"]]**
168 168  
169 -\\
170 170  
171 -\\
172 172  
173 -**
174 -**
164 +
175 175  {{/layout-cell}}
176 176  {{/layout-section}}
177 177  
... ... @@ -185,13 +185,13 @@
185 185  
186 186  (% class="code" %)
187 187  (((
188 - cd /gpfs/local/shared/ChopperControl
178 + cd /gpfs/local/shared/ChopperControl          
189 189  )))
190 190  
191 191  (% class="code" %)
192 192  (((
193 193   python chopperControl.py
194 -\\\\
184 +\\
195 195  )))
196 196  {{/layout-cell}}
197 197  {{/layout-section}}
... ... @@ -204,18 +204,17 @@
204 204  {{layout-cell}}
205 205  **Chopper → no Chopper**
206 206  
207 -* decrease frequency by max. 10 Hz steps\\
197 +* decrease frequency by max. 10 Hz steps
208 208  ** click arrow down 10x, "Set", arrow down 10x, "Set", ....
209 209  * at frequency = 10 Hz, click "Stop"
210 -* set "Hole" to <none> (% style="color: rgb(255,0,0);" %)//Takes some minutes//
200 +* set "Hole" to <none> (% style="color:#ff0000" %)//Takes some minutes//
211 211  
212 -(% style="color: rgb(0,0,0);" %)**no Chopper → Chopper**
202 +(% style="color:#000000" %)**no Chopper → Chopper**
213 213  
214 -* (% style="color: rgb(0,0,0);" %)set "Hole" to <Inner hole (largest)> (% style="color: rgb(255,0,0);" %)//Takes some minutes//
215 -* (% style="color: rgb(0,0,0);" %)increase frequency by max. 10 Hz steps
216 -** (% style="color: rgb(0,0,0);" %)click arrow up 10x, "Set", arrow up 10x, "Set", .....
217 -* (% style="color: rgb(0,0,0);" %)//usually 130 Hz, results in exposure time of 3.7 ms (technical exposure will always be 7.7 ms, since max. frame rate is 130 Hz
218 -//
204 +* (% style="color:#000000" %)set "Hole" to <Inner hole (largest)> (% style="color:#ff0000" %)//Takes some minutes//
205 +* (% style="color:#000000" %)increase frequency by max. 10 Hz steps
206 +** (% style="color:#000000" %)click arrow up 10x, "Set", arrow up 10x, "Set", .....
207 +* (% style="color:#000000" %)//usually 130 Hz, results in exposure time of 3.7 ms (technical exposure will always be 7.7 ms, since max. frame rate is 130 Hz//
219 219  {{/layout-cell}}
220 220  {{/layout-section}}
221 221  
... ... @@ -235,7 +235,7 @@
235 235  
236 236   //>>usually the terminal is open, just repeat previous command with arrow up, enter and type in the password//
237 237  
238 -\\
227 +
239 239  {{/layout-cell}}
240 240  {{/layout-section}}
241 241  
... ... @@ -243,22 +243,19 @@
243 243  {{layout-cell}}
244 244  **Check list:**
245 245  
246 -(% style="color: rgb(255,0,0);" %)//always after entering the hutch!//
235 +(% style="color:#ff0000" %)//always after entering the hutch!//
247 247  
248 -(% style="color: rgb(0,0,0);" %)Click "Run" icon in kamzik.
237 +(% style="color:#000000" %)Click "Run" icon in kamzik.
249 249  
250 -* Put detector distance 200 in //prepare_for_measurement (      )
251 -//
252 -* Click "<Call>" next to //prepare_for_measurement
253 -//This will:\\
239 +* Put detector distance 200 in //prepare_for_measurement (      )//
240 +* Click "<Call>" next to //prepare_for_measurement//
241 +This will:
254 254  ** drive the collimator in //IN// position
255 255  ** drive detector to 200 mm
256 256  
257 -\\
258 258  
259 -* Open beam shutter BS1 in //CrystalControl //or// P11 XRayFeed //
246 +* Open beam shutter BS1 in //CrystalControl //or// P11 XRayFeed //
260 260  
261 -\\
262 262  
263 263  //**Set run parameters and go 🏃💥**//
264 264  
... ... @@ -266,9 +266,8 @@
266 266  1. Enter the protein ID of your crystal sample
267 267  1. Match exposure time with chopper setting or set desired exposure time
268 268  1. Check trigger setting and match to chopper setting
269 -no chopper → external triffer //off
270 -// chopper → external trigger// on
271 -//
255 +no chopper → external triffer //off//
256 + chopper → external trigger// on//
272 272  1. To start the run, click "<Call>" next to //start_run (exposure    )//
273 273  {{/layout-cell}}
274 274  
... ... @@ -279,7 +279,7 @@
279 279  
280 280  {{layout-section ac:type="single"}}
281 281  {{layout-cell}}
282 -(% class="relative-table wrapped" style="width: 55.9569%;" %)
267 +(% class="relative-table wrapped" style="width:55.9569%" %)
283 283  |=(((
284 284  Parameter
285 285  )))|=(((
... ... @@ -301,7 +301,7 @@
301 301  )))|(((
302 302  exposure time of Eiger
303 303  
304 -\\
289 +
305 305  )))|(% colspan="1" %)(((
306 306  with chopper         → opening time displayed in chopper GUI
307 307  no chopper            → 7.7 ms
... ... @@ -313,7 +313,7 @@
313 313  )))|(((
314 314  changes automatically
315 315  )))|(% colspan="1" %)(((
316 -\\
301 +
317 317  )))
318 318  |(% colspan="1" %)(((
319 319  Detector distance
... ... @@ -320,7 +320,7 @@
320 320  )))|(% colspan="1" %)(((
321 321  current detector postion readout
322 322  )))|(% colspan="1" %)(((
323 -\\
308 +
324 324  )))
325 325  |(% colspan="1" %)(((
326 326  protein ID
... ... @@ -327,7 +327,7 @@
327 327  )))|(% colspan="1" %)(((
328 328  protein ID of crystal sample (see AMARCORD)
329 329  )))|(% colspan="1" %)(((
330 -\\
315 +
331 331  )))
332 332  
333 333  ----
... ... @@ -348,15 +348,11 @@
348 348  {{/layout-cell}}
349 349  
350 350  {{layout-cell}}
351 -\\
336 +
352 352  
353 -\\
354 354  
355 -\\
356 356  
357 -\\
358 358  
359 -\\
360 360  
361 361  [[image:attach:safety_kamzik.PNG]]
362 362  {{/layout-cell}}
... ... @@ -375,7 +375,6 @@
375 375  * motor connected (icon green)?
376 376  )))
377 377  
378 -\\
379 379  
380 380  * (((
381 381  ==== {{id name="14.SerialCrystallography:CFELTapeDrive-Nozzlemotorsarenotresponding"/}}Nozzle motors are not responding ====
... ... @@ -385,7 +385,6 @@
385 385  * motor limits reached? → position displayed in red
386 386  )))
387 387  
388 -\\
389 389  
390 390  * (((
391 391  ==== {{id name="14.SerialCrystallography:CFELTapeDrive-Sampleflowisnotrespondingtohigherpressure"/}}Sample flow is not responding to higher pressure ====
... ... @@ -395,7 +395,6 @@
395 395  * Nozzle or lines clogged? → flush with water, change nozzle
396 396  )))
397 397  
398 -\\
399 399  
400 400  * (((
401 401  ==== {{id name="14.SerialCrystallography:CFELTapeDrive-Simplonfilewriter/eigerstreamissues"/}}Simplon filewriter/eiger stream issues ====
... ... @@ -403,12 +403,15 @@
403 403  * First, restart kamzik (server+client) and see if this solves the issue
404 404  * If not, restart the Simplon server (in Astor → haspp11exp03, right click on Simplon/EH.1 and restart)
405 405  * (((
406 -If nothing works, kill the Simplon process and then start the server again:* (((
384 +If nothing works, kill the Simplon process and then start the server again:*
385 +
386 +(((
407 407  (% class="code" %)
408 408  (((
409 409  ssh haspp11exp03
410 410  )))
411 411  )))
392 +
412 412  * (((
413 413  (% class="code" %)
414 414  (((
... ... @@ -429,7 +429,7 @@
429 429  
430 430  {{layout-section ac:type="single"}}
431 431  {{layout-cell}}
432 -\\
413 +
433 433  {{/layout-cell}}
434 434  {{/layout-section}}
435 435  {{/layout}}