Title
... ... @@ -1,1 +1,1 @@
1 -10. Serial Crystallography: CFEL TapeDrive
1 +08. Serial Crystallography: CFEL TapeDrive
Content
... ... @@ -24,7 +24,7 @@
24 24  
25 25  == {{id name="14.SerialCrystallography:CFELTapeDrive-14.2ConnectingsampletoElveFlow,sampleexchange"/}}**14.2 Connecting sample to ElveFlow, sample exchange** ==
26 26  
27 -*
27 +*
28 28  ** ==
29 29  
30 30  After filtration, screw the ElveFlow connector to your Eppendorf or Falcon tube. Make sure that the yellow-ish line goes to the bottom of your tube. You can adjust the length by unscrewing the blue fitting on top and pushing/pulling the yellow-ish line. **Don't forget to tigthen the blue fitting afterwards!**
... ... @@ -46,7 +46,7 @@
46 46  
47 47  == {{id name="14.SerialCrystallography:CFELTapeDrive-14.3TapeDriveoperation"/}}**14.3 TapeDrive operation** ==
48 48  
49 -*
49 +*
50 50  ** ==
51 51  
52 52  The TapeDrive is controlled through kamzik. To open kamzik, start the server. A terminal will open and shortly after a small window.
... ... @@ -116,7 +116,6 @@
116 116  be careful with -z, tape could be pushed out of focus
117 117  )))
118 118  
119 -
120 120  
121 121  )))
122 122  * [[image:attach:kamzik_nozzle.PNG||height="400"]]
... ... @@ -125,7 +125,6 @@
125 125  
126 126  * The chiller circulates fluid through the gas nozzle.
127 127  
128 -
129 129  ==== {{id name="14.SerialCrystallography:CFELTapeDrive-Gasnozzle"/}}**Gas nozzle** ====
130 130  
131 131  
... ... @@ -242,10 +242,8 @@
242 242  ** drive the collimator in //IN// position
243 243  ** drive detector to 200 mm
244 244  
245 -
246 246  * Open beam shutter BS1 in //CrystalControl //or// P11 XRayFeed //
247 247  
248 -
249 249  //**Set run parameters and go 🏃💥**//
250 250  
251 251  1. Set your desired No. of frames
... ... @@ -356,7 +356,6 @@
356 356  * motor connected (icon green)?
357 357  )))
358 358  
359 -
360 360  * (((
361 361  ==== {{id name="14.SerialCrystallography:CFELTapeDrive-Nozzlemotorsarenotresponding"/}}Nozzle motors are not responding ====
362 362  
... ... @@ -365,7 +365,6 @@
365 365  * motor limits reached? → position displayed in red
366 366  )))
367 367  
368 -
369 369  * (((
370 370  ==== {{id name="14.SerialCrystallography:CFELTapeDrive-Sampleflowisnotrespondingtohigherpressure"/}}Sample flow is not responding to higher pressure ====
371 371  
... ... @@ -374,7 +374,6 @@
374 374  * Nozzle or lines clogged? → flush with water, change nozzle
375 375  )))
376 376  
377 -
378 378  * (((
379 379  ==== {{id name="14.SerialCrystallography:CFELTapeDrive-Simplonfilewriter/eigerstreamissues"/}}Simplon filewriter/eiger stream issues ====
380 380