Docu of the FLASH2 camera trigger system

Version 3.1 by sndueste on 2022/12/21 15:29
Warning: For security reasons, the document is displayed in restricted mode as it is not the current version. There may be differences and errors due to this.

The beamline cameras and the camera triggers on the User patch panels are triggered by different  MTCA based triggers that are distributed using several Trigger boxes ( with 1x 5V input and 10x5V outputs)

(there is a 12 V input option which one can set with jumpers. contact Marc Temme - he organized the design and building by DESY electronic workshop)


image2022-12-21_14-20-49.png


  • The trigger for the tunnel is based on a trigger from a DIAG crate  FLASH.DIAG/TIMER/F2DIAG13/FRONT.TRG2  . this trigger is connected to a Beckhoff IO box ( flash2tpdiagio ,  192.168.144.167 ). from there the trigger signals are transported in the tunnel with the cables with many wires (24 ?) connected to the "Klemmblöcke"

  • the FL20 trigger for the first part of the FLASH hall ( Det3, Det4, FL21, screens at M3, ...) is taken from the GMD crate FLASH.FEL/TIMER/XGMD2/FRONT.TRG4. this is also connected to a Beckhoff IO box ( flash2hpdiagio ,   192.168.144.168 ). from there the trigger signals are transported in the tunnel with the cables with many wires (24 ?) connected to the "Klemmblöcke"
  • The camera trigger for FL24 and FL23 is taken from the exp1 crate in Rack 41  FLASH.FEL/TIMER/FL2EXP1.1/RTM.TRG9  . The trigger is brought to the FL24 patch panel and to an amplifier box. from there we feed the cameras and more amplifier boxes. see image below
  • The camera trigger for FL26 is taken from the exp1 crate in Rack 41  FLASH.FEL/TIMER/FL2EXP1/RTM.TRG5   . The trigger is brought to the FL26 patch panel and to an amplifier box feeding mainly the FL26 patch panel and a second one next to it which triggers the beamline cameras at the REMI