Changes for page Gotthard related stuff

Last modified by sndueste on 2023/09/25 09:36

From version 12.1
edited by sndueste
on 2023/07/13 16:42
Change comment: There is no comment for this version
To version 14.1
edited by sndueste
on 2023/09/08 09:10
Change comment: There is no comment for this version

Summary

Details

Page properties
Content
... ... @@ -5,6 +5,7 @@
5 5  * [[Data sheet of a macro objective>>attach:MP-E_65mm_F2.8_1-5x_Instruction_DE.pdf]]
6 6  * [[FLASH-GH-Meeting_10.10.2014.pdf by Steffen>>attach:FLASH-GH-Meeting_10.10.2014.pdf]]
7 7  * [[attach:READ_ME_GOTTHARD_detector.pdf]]: READ_ME_GOTTHARD_detector.pdf (Server information is outdated, see below)
8 +* some investigation related to read out noise for different exposure times [[attach:GOTTHARD noise evaluation and exposure time dependence (002).pdf]]
8 8  
9 9  \\
10 10  
... ... @@ -54,7 +54,7 @@
54 54  
55 55  (% style="list-style-type: square;" %)
56 56  * Do not change the delay in the GOTTHARD GUI. The Alignment to the FEL pulses are done by adjusting the FLASH trigger: FL2EXP1/RTM.TRG6, it can be opened from the GOTTHARD GUI or from FLASH2 → Beamlines → FL26 → Exp. control → Patch Panel26-2 → MTCA Trigger 2
57 -* Exposure range is **100ns <  Exposure < ~~1/3 * Frame periode**
58 +* Exposure range is 2**00ns <  Exposure < ~~1/3 * Frame periode**
58 58  * The periode is:** 9968ns for 100kHz, 4984 for 200kHz, 1993 for 500kHz**
59 59  * The number of** frames must be lower than 125, for stable operation 120**
60 60  * Number of Shots: Number of trigger events before self stopping, select large number for permanent operation