Changes for page Gotthard related stuff

Last modified by sndueste on 2023/09/25 09:36

From version 14.1
edited by sndueste
on 2023/09/08 09:10
Change comment: There is no comment for this version
To version 11.1
edited by palutke
on 2023/06/24 16:24
Change comment: There is no comment for this version

Summary

Details

Page properties
Author
... ... @@ -1,1 +1,1 @@
1 -XWiki.sndueste
1 +XWiki.palutke
Content
... ... @@ -5,7 +5,6 @@
5 5  * [[Data sheet of a macro objective>>attach:MP-E_65mm_F2.8_1-5x_Instruction_DE.pdf]]
6 6  * [[FLASH-GH-Meeting_10.10.2014.pdf by Steffen>>attach:FLASH-GH-Meeting_10.10.2014.pdf]]
7 7  * [[attach:READ_ME_GOTTHARD_detector.pdf]]: READ_ME_GOTTHARD_detector.pdf (Server information is outdated, see below)
8 -* some investigation related to read out noise for different exposure times [[attach:GOTTHARD noise evaluation and exposure time dependence (002).pdf]]
9 9  
10 10  \\
11 11  
... ... @@ -55,9 +55,9 @@
55 55  
56 56  (% style="list-style-type: square;" %)
57 57  * Do not change the delay in the GOTTHARD GUI. The Alignment to the FEL pulses are done by adjusting the FLASH trigger: FL2EXP1/RTM.TRG6, it can be opened from the GOTTHARD GUI or from FLASH2 → Beamlines → FL26 → Exp. control → Patch Panel26-2 → MTCA Trigger 2
58 -* Exposure range is 2**00ns <  Exposure < ~~1/3 * Frame periode**
59 -* The periode is:** 9968ns for 100kHz, 4984 for 200kHz, 1993 for 500kHz**
60 -* The number of** frames must be lower than 125, for stable operation 120**
57 +* Exposure range is 100ns <  Exposure < ~~1/3 * Frame periode
58 +* The periode is: 9968ns for 100kHz, 4984 for 200kHz, 1993 for 500kHz
59 +* The number of frames must be lower than 125, for stable operation 120
61 61  * Number of Shots: Number of trigger events before self stopping, select large number for permanent operation
62 62  
63 63  == Modules and their exchange ==
... ... @@ -95,8 +95,6 @@
95 95  )))
96 96  |(((
97 97  hasdslabgh
98 -
99 -"gotthard dgt01"
100 100  )))|(((
101 101  working
102 102  )))|(((