... ... @@ -1,5 +2,9 @@
1 -== {{id name="Loadframe-Overview"/}}Overview ==
2 2  
2 +
3 +{{toc/}}
4 +
5 += {{id name="Loadframe-Overview"/}}Overview =
6 +
3 3  [[image:attach:DSC_0006.JPG||width="400"]]
4 4  
5 5  * Forces up to 1 kN (both pushing or pulling)
... ... @@ -6,7 +6,7 @@
6 6  * Load cells 20N, 50N, 100N, 200N, 500N and 1000N available
7 7  * Fully integrated into control and logging system
8 8  
9 -== {{id name="Loadframe-Actuator"/}}Actuator ==
13 += {{id name="Loadframe-Actuator"/}}Actuator =
10 10  
11 11  [[image:attach:DSC_0020.JPG||height="250"]][[image:attach:DSC_0021.JPG||height="250"]]
12 12  
... ... @@ -50,7 +50,7 @@
50 50  )))
51 51  )))
52 52  
53 -== {{id name="Loadframe-Loadcells"/}}Load cells ==
57 += {{id name="Loadframe-Loadcells"/}}Load cells =
54 54  
55 55  Two types of load cells are available with different diameters with the following force range
56 56  
... ... @@ -102,19 +102,20 @@
102 102  thread: M4 (max len 5 mm)
103 103  )))
104 104  
105 -== {{id name="Loadframe-Loadframeassembly"/}}Load frame assembly ==
109 += {{id name="Loadframe-Loadframeassembly"/}}Load frame assembly =
106 106  
107 -Assembing the load frame can be brikne down into these steps:
111 +Considering different applications (high forces, low forces, pushing or pulling) the assembly follows slightly different paths. Generally, assembling the load frame can be broken down into these steps:
108 108  
109 -* Install a base plate on the rotation stage (use spacers)
110 -* Mount load cell onS base plate with desired pin or clamp
111 -* Mount sample tube
113 +* Install a base plate on the rotation stage (use spacers, decide which plate to use)
114 +* Mount load cell on base plate with desired pin or clamp (depending on push or pull experiments)
115 +* Mount sample tube (choose the tube type according to your application)
112 112  * Mount pusher
113 113  
114 -=== {{id name="Loadframe-Largeloadcellinstallation"/}}Large load cell installation ===
118 +== {{id name="Loadframe-Largeloadcellinstallation"/}}Large load cell installation ==
115 115  
116 -For the large load cell a specific base plate has to be used (see image)
120 +For the large load cell a specific base plate has to be used
117 117  
122 +(% class="wrapped" %)
118 118  |(% style="text-align: center;" %)(((
119 119  (% class="content-wrapper" %)
120 120  (((
... ... @@ -132,8 +132,217 @@
132 132  Load cell with pin used for pulling
133 133  )))
134 134  
135 -== {{id name="Loadframe-Reference"/}}Reference ==
140 +Mount bridge and guide - there are two different types of tube, so assembly depends on your choice of tube:
136 136  
142 +=== {{id name="Loadframe-Tubewiththreadring"/}}Tube with thread ring ===
143 +
144 +(% class="wrapped" %)
145 +|(% style="text-align: center;" %)(((
146 +(% class="content-wrapper" %)
147 +(((
148 +[[image:attach:DSC_0034.JPG||height="250"]]
149 +)))
150 +)))|(% style="text-align: center;" %)(((
151 +(% class="content-wrapper" %)
152 +(((
153 +[[image:attach:DSC_0035.JPG||height="250"]]
154 +)))
155 +)))|(% style="text-align: center;" %)(((
156 +(% class="content-wrapper" %)
157 +(((
158 +[[image:attach:DSC_0036.JPG||height="250"]]
159 +)))
160 +)))
161 +|(% style="text-align: center;" %)(((
162 +Bridge
163 +)))|(% style="text-align: center;" %)(((
164 +Brigde and pin guide
165 +)))|(% style="text-align: center;" %)(((
166 +Tube with thread ring
167 +)))
168 +
169 +=== {{id name="Loadframe-Tubewithinternalthread"/}}Tube with internal thread ===
170 +
171 +(% class="wrapped" %)
172 +|(% style="text-align: center;" %)(((
173 +(% class="content-wrapper" %)
174 +(((
175 +[[image:attach:DSC_0034.JPG||height="250"]]
176 +)))
177 +)))|(% style="text-align: center;" %)(((
178 +(% class="content-wrapper" %)
179 +(((
180 +[[image:attach:DSC_0037.JPG||thumbnail="true" height="250"]]
181 +)))
182 +)))|(% style="text-align: center;" %)(((
183 +(% class="content-wrapper" %)
184 +(((
185 +[[image:attach:DSC_0038.JPG||height="250"]]
186 +)))
187 +)))
188 +|(% style="text-align: center;" %)(((
189 +Bridge
190 +)))|(% style="text-align: center;" %)(((
191 +Bridge and pin guide
192 +)))|(% style="text-align: center;" %)(((
193 +Tube with internal thread
194 +)))
195 +
196 +== {{id name="Loadframe-Smallloadcellinstallation"/}}Small load cell installation ==
197 +
198 +For the small load cell a specific base plate has to be used
199 +
200 +(% class="wrapped" %)
201 +|(% style="text-align: center;" %)(((
202 +(% class="content-wrapper" %)
203 +(((
204 +[[image:attach:DSC_0039.JPG||height="250"]]
205 +)))
206 +)))|(% style="text-align: center;" %)(((
207 +(% class="content-wrapper" %)
208 +(((
209 +[[image:attach:DSC_0040.JPG||height="250"]]
210 +)))
211 +)))
212 +|(% style="text-align: center;" %)(((
213 +Base plate with load cell
214 +)))|(% style="text-align: center;" %)(((
215 +Load cell with pin used for pulling
216 +)))
217 +
218 +Mount pin guide and bridge - there are two different types of tube, so assembly depends on your choice of tube:
219 +
220 +=== {{id name="Loadframe-Tubewiththreadring"/}}Tube with thread ring ===
221 +
222 +(% class="wrapped" %)
223 +|(((
224 +(% class="content-wrapper" %)
225 +(((
226 +[[image:attach:DSC_0041.JPG||height="250"]]
227 +)))
228 +)))|(((
229 +(% class="content-wrapper" %)
230 +(((
231 +[[image:attach:DSC_0042.JPG||height="250"]]
232 +)))
233 +)))
234 +|(((
235 +Pin guide, no bridge necessary
236 +)))|(((
237 +Tube with thread ring
238 +)))
239 +
240 +=== {{id name="Loadframe-Tubewithinternalthread"/}}Tube with internal thread ===
241 +
242 +(% class="wrapped" %)
243 +|(% style="text-align: center;" %)(((
244 +(% class="content-wrapper" %)
245 +(((
246 +[[image:attach:DSC_0043.JPG||height="250"]]
247 +)))
248 +)))|(% style="text-align: center;" %)(((
249 +(% class="content-wrapper" %)
250 +(((
251 +[[image:attach:DSC_0044.JPG||height="250"]]
252 +)))
253 +)))|(% style="text-align: center;" %)(((
254 +(% class="content-wrapper" %)
255 +(((
256 +[[image:attach:DSC_0045.JPG||height="250"]]
257 +)))
258 +)))|(% style="text-align: center;" %)(((
259 +(% class="content-wrapper" %)
260 +(((
261 +[[image:attach:DSC_0046.JPG||height="250"]]
262 +)))
263 +)))
264 +|(% style="text-align: center;" %)(((
265 +Pin guide
266 +)))|(% style="text-align: center;" %)(((
267 +Bridge
268 +)))|(% style="text-align: center;" %)(((
269 +Optionally: insert push pin
270 +)))|(% style="text-align: center;" %)(((
271 +Tube with internal thread
272 +)))
273 +
274 += {{id name="Loadframe-Tubes"/}}Tubes =
275 +
276 +{{info title="Tubes"}}
277 +There are different tubes available at the beamline. Should the tube be unsuitable for your sample, a new tube can be manufactured. Please contact the beamline staff well in advance of your beamtime.
278 +{{/info}}
279 +
280 +(% class="wrapped" %)
281 +|(% style="text-align: center;" %)(((
282 +(% class="content-wrapper" %)
283 +(((
284 +[[image:attach:DSC_0028.JPG||height="250"]]
285 +)))
286 +)))
287 +|(% style="text-align: center;" %)(((
288 +Different Tubes with lengths of ~~80 mm and 140 mm
289 +)))
290 +
291 += {{id name="Loadframe-Samplemounting"/}}Sample mounting =
292 +
293 +{{info title="Pins and clamps"}}
294 +Depending on your application, sample installation may vary. There are a selection of pushing pins and pulling clamps available at the beamline which will be presented in this section. Nevertheless these pins or clamps may be unsuitable for your sample. In this case a suitable pin or clamp can be manufactured. Please contact the beamline staff well in advance of your beamtime.
295 +{{/info}}
296 +
297 +(% class="wrapped" %)
298 +|=(% colspan="2" %)(((
299 +Pushing
300 +)))|=(% colspan="2" %)(((
301 +Pulling
302 +)))
303 +|(((
304 +Bottom push pin
305 +)))|(((
306 +(% class="content-wrapper" %)
307 +(((
308 +[[image:attach:DSC_0045.JPG||height="250"]]
309 +)))
310 +)))|(((
311 +Bottom pull thread
312 +)))|(((
313 +(% class="content-wrapper" %)
314 +(((
315 +[[image:attach:DSC_0041.JPG||height="250"]]
316 +)))
317 +)))
318 +|(((
319 +Top push pin
320 +)))|(((
321 +(% class="content-wrapper" %)
322 +(((
323 +[[image:attach:DSC_0047.JPG||height="250"]]
324 +)))
325 +)))|(((
326 +Principle of pull installation
327 +)))|(((
328 +(% class="content-wrapper" %)
329 +(((
330 +[[image:attach:DSC_0049.JPG||height="250"]]
331 +)))
332 +)))
333 +|(((
334 +Top push pin
335 +)))|(((
336 +(% class="content-wrapper" %)
337 +(((
338 +[[image:attach:DSC_0048.JPG||height="250"]]
339 +)))
340 +)))|(((
341 +Pull clamps for low forces
342 +)))|(((
343 +(% class="content-wrapper" %)
344 +(((
345 +[[image:attach:DSC_0050.JPG||thumbnail="true" height="250"]]
346 +)))
347 +)))
348 +
349 += {{id name="Loadframe-Reference"/}}Reference =
350 +
137 137  Moosmann et al. "A load frame for in situ tomography at PETRA III", Proc. SPIE 11113, Developments in X-Ray Tomography XII, 1111318 (2019); [[https:~~/~~/doi.org/10.1117/12.2530445>>url:https://doi.org/10.1117/12.2530445||rel="nofollow" shape="rect"]]